AIRYX · ICAD series · direct HONO and NO₂ measurement

ICAD-HONO/NO₂-210L

A high-sensitivity in-situ analyser for simultaneous, direct determination of nitrous acid and nitrogen dioxide. Both species are resolved spectroscopically — without a converter, reagents, sample drying or external calibration-gas standards. It is designed for atmospheric chemistry research and measurements in laboratories and research chambers.

AIRYX ICAD-HONO/NO2-210L analyserICAD-HONO/NO₂-210L
0.08 ppbHONO detection limit at 60 s
0.16 ppbNO₂ detection limit at 60 s
1 sminimum averaging time
up to 3 kmeffective absorption path
Fully optical measurement

HONO and NO₂ resolved in one spectrum

Patented Iterative Cavity DOAS spectroscopy uses characteristic absorption structures of both gases around 350–390 nm. The resonator creates an optical path of up to 3 km, enabling very low concentrations to be measured with a response suitable for dynamic research campaigns.

No conversion or sample conditioning

direct, simultaneous HONO and NO₂ measurement
single High Grade optical resonator cell
no converter or chemical reagents
no sample drying required
automatic zero measurements to limit drift
optional NDIR CO₂ sensor
ICAD-HONO/NO2-210L optical measurement-path diagram
High Grade — L version

Precision for atmospheric chemistry

An extended resonator with High Grade mirrors increases sensitivity, while active Peltier cooling stabilises LED temperature to better than 0.1°C. Longer averaging improves precision without sacrificing spectral selectivity.

HONO precision versus averaging time

Precision follows measurement time

At 60 s, 1σ precision is 0.04 ppb for HONO and 0.08 ppb for NO₂. The standard delivery setting is 30 s and can be changed in the Web UI.

ICAD-HONO/NO2-210L interface on a tablet

Both species visible at the same time

The browser interface displays HONO and NO₂ traces, measurement settings and the System Health table. No separate application is required.

Stability and simple operation

Ready for extended measurement campaigns

The system is insensitive to typical vibration and changes in absolute light-source intensity. On-board diagnostics monitor flow, supply voltage, temperature and other critical instrument parameters.

◎

No calibration gas

HONO and NO₂ concentrations are derived from molecular absorption cross sections, while automatic zero checks ensure long-term stability.

↯

Fast response

Response is approximately 1 s at a 2 l/min flow and remains below 5 s at 1.5 l/min.

⌁

Remote access

Results, configuration and diagnostics are available over LAN or Wi-Fi from a computer, tablet or phone.

Instrument technical data

ICAD-HONO/NO₂-210L specification

Parameters refer to the high-sensitivity 19″ rack version. Custom configurations and ranges are available; a reduced range can further improve precision and the detection limit.

Analytical performance
Measured speciesHONO and NO₂ — direct, simultaneous in-situ spectroscopic measurement
HONO range0–500 ppb
NO₂ range0–2000 ppb
Detection limit, 2σ at 60 sHONO: 0.08 ppb; NO₂: 0.16 ppb
Precision, 1σ at 1 sHONO: 0.3 ppb; NO₂: 0.6 ppb
Precision, 1σ at 60 sHONO: 0.04 ppb; NO₂: 0.08 ppb
Time resolution1 s minimum averaging; 30 s standard delivery setting
Response time, 10–90%approximately 1 s at 2 l/min; below 5 s at 1.5 l/min; flow is adjustable
Zero driftHONO: less than 0.2 ppb/month; NO₂: less than 0.3 ppb/month; regular automatic zero measurements further reduce noise and drift
Temperature sensitivityless than 0.01 ppb/°C
Spectral rangeapproximately 350–390 nm
Absorption pathup to 3 km in the optical resonator
Measurement cellenhanced-sensitivity resonator with High Grade mirrors
Light-source stabilisationactive Peltier LED cooling; temperature stability better than 0.1°C
Other detectable gasesCO₂ with an optional NDIR sensor
Cross-sensitivityno significant spectroscopic cross-sensitivity to, among others, carbon oxides, formaldehyde, ozone, methane, hydrogen, sulphide, SO₂, chlorine, ClO₂, HCN, HCl, phosphine, ammonia, acetylene, nitromethane, ethylene, ethanol, methyl mercaptan or ethyl mercaptan
Sample, zero and calibration
Sample flow1–2 l/min
Sample preparationsample drying is not required; a PTFE aerosol filter is used at the inlet
HONO and NO₂ calibrationcalibration gases are not required; concentrations are quantified from literature absorption data
Optical-path characterisationoptional helium every 1–2 years or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version
Zero-air systemlarge zero-air cartridge included; regular- and high-capacity systems are available
Consumable gases and chemicalsnone required during normal operation
Construction and operating conditions
Enclosure19″ rack, 3U high, with a front-panel OLED display
Dimensions43.8 × 13.3 × 66.5 cm (W × H × D)
Weightless than 12 kg, depending on configuration
Operating temperature-10 to +40°C
Mechanical stabilityinsensitive to typical vibration
Power12 V DC; external ACC-PWR-8A power supply required
Power consumptiontypically less than 40 W at 12 V
Start-up timetypically less than 1 min
Mobile usethe rack instrument can be used in suitably mounted mobile installations; it is not the IP64 M field version
Control, communication and equipment
Processing unitembedded industrial computer with data-analysis and measurement software
User interfaceWeb UI with HONO and NO₂ time series, span-calibration adjustment, System Health table and on-board diagnostics
Data interfacesLAN, Wi-Fi, RS232, Machine-to-Machine, OPC UA, Bayern-Hessen protocol and analogue voltage/current outputs
GPSoptional IC-ACC-GPSEXT module for geolocation and time synchronisation
Included equipmentlarge zero-air cartridge, PTFE aerosol filter holder with filter, 5 m PFA tubing and 3 m LAN cable
Required itemsexternal ACC-PWR-8A power supply; ACC-WIFIUSB module when Wi-Fi connectivity is required
Options and accessoriesoperator tablet, GPS, ICAD cooling, integrated NDIR CO₂ sensor, PTFE filters, regular- or high-capacity zero-air system, maintenance set and IC-SPARE-SET-HONO spare-part set
PatentsEP3329251; D502016003239.4; GB3329251; FR3329251; US10473583B2; ZL201680057099.6; DE102015000423; EP325928; D502016014735.3; GB3325928; FR3325928
HONO/NO₂ applications

A direct view of atmospheric chemistry

Simultaneous HONO and NO₂ measurement reveals processes that a conventional NOx-only analyser cannot resolve.

Atmospheric HONO researchnight-time chemistry, HONO photolysis, OH-radical sources and photochemical-model validation
Air-quality monitoringurban-process studies, source attribution and parallel observation of HONO and NO₂
Laboratories and research chamberssmog-chamber experiments, method comparisons, reaction kinetics and reference work
Mobile field campaignsfast measurements on vehicles and ships when the rack instrument is correctly installed