Precision follows measurement time
At 60 s, 1σ precision is 0.04 ppb for HONO and 0.08 ppb for NO₂. The standard delivery setting is 30 s and can be changed in the Web UI.
A high-sensitivity in-situ analyser for simultaneous, direct determination of nitrous acid and nitrogen dioxide. Both species are resolved spectroscopically — without a converter, reagents, sample drying or external calibration-gas standards. It is designed for atmospheric chemistry research and measurements in laboratories and research chambers.
ICAD-HONO/NO₂-210L
Patented Iterative Cavity DOAS spectroscopy uses characteristic absorption structures of both gases around 350–390 nm. The resonator creates an optical path of up to 3 km, enabling very low concentrations to be measured with a response suitable for dynamic research campaigns.
An extended resonator with High Grade mirrors increases sensitivity, while active Peltier cooling stabilises LED temperature to better than 0.1°C. Longer averaging improves precision without sacrificing spectral selectivity.
At 60 s, 1σ precision is 0.04 ppb for HONO and 0.08 ppb for NO₂. The standard delivery setting is 30 s and can be changed in the Web UI.
The browser interface displays HONO and NO₂ traces, measurement settings and the System Health table. No separate application is required.
The system is insensitive to typical vibration and changes in absolute light-source intensity. On-board diagnostics monitor flow, supply voltage, temperature and other critical instrument parameters.
HONO and NO₂ concentrations are derived from molecular absorption cross sections, while automatic zero checks ensure long-term stability.
Response is approximately 1 s at a 2 l/min flow and remains below 5 s at 1.5 l/min.
Results, configuration and diagnostics are available over LAN or Wi-Fi from a computer, tablet or phone.
Parameters refer to the high-sensitivity 19″ rack version. Custom configurations and ranges are available; a reduced range can further improve precision and the detection limit.
| Analytical performance | |
|---|---|
| Measured species | HONO and NO₂ — direct, simultaneous in-situ spectroscopic measurement |
| HONO range | 0–500 ppb |
| NO₂ range | 0–2000 ppb |
| Detection limit, 2σ at 60 s | HONO: 0.08 ppb; NO₂: 0.16 ppb |
| Precision, 1σ at 1 s | HONO: 0.3 ppb; NO₂: 0.6 ppb |
| Precision, 1σ at 60 s | HONO: 0.04 ppb; NO₂: 0.08 ppb |
| Time resolution | 1 s minimum averaging; 30 s standard delivery setting |
| Response time, 10–90% | approximately 1 s at 2 l/min; below 5 s at 1.5 l/min; flow is adjustable |
| Zero drift | HONO: less than 0.2 ppb/month; NO₂: less than 0.3 ppb/month; regular automatic zero measurements further reduce noise and drift |
| Temperature sensitivity | less than 0.01 ppb/°C |
| Spectral range | approximately 350–390 nm |
| Absorption path | up to 3 km in the optical resonator |
| Measurement cell | enhanced-sensitivity resonator with High Grade mirrors |
| Light-source stabilisation | active Peltier LED cooling; temperature stability better than 0.1°C |
| Other detectable gases | CO₂ with an optional NDIR sensor |
| Cross-sensitivity | no significant spectroscopic cross-sensitivity to, among others, carbon oxides, formaldehyde, ozone, methane, hydrogen, sulphide, SO₂, chlorine, ClO₂, HCN, HCl, phosphine, ammonia, acetylene, nitromethane, ethylene, ethanol, methyl mercaptan or ethyl mercaptan |
| Sample, zero and calibration | |
| Sample flow | 1–2 l/min |
| Sample preparation | sample drying is not required; a PTFE aerosol filter is used at the inlet |
| HONO and NO₂ calibration | calibration gases are not required; concentrations are quantified from literature absorption data |
| Optical-path characterisation | optional helium every 1–2 years or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version |
| Zero-air system | large zero-air cartridge included; regular- and high-capacity systems are available |
| Consumable gases and chemicals | none required during normal operation |
| Construction and operating conditions | |
| Enclosure | 19″ rack, 3U high, with a front-panel OLED display |
| Dimensions | 43.8 × 13.3 × 66.5 cm (W × H × D) |
| Weight | less than 12 kg, depending on configuration |
| Operating temperature | -10 to +40°C |
| Mechanical stability | insensitive to typical vibration |
| Power | 12 V DC; external ACC-PWR-8A power supply required |
| Power consumption | typically less than 40 W at 12 V |
| Start-up time | typically less than 1 min |
| Mobile use | the rack instrument can be used in suitably mounted mobile installations; it is not the IP64 M field version |
| Control, communication and equipment | |
| Processing unit | embedded industrial computer with data-analysis and measurement software |
| User interface | Web UI with HONO and NO₂ time series, span-calibration adjustment, System Health table and on-board diagnostics |
| Data interfaces | LAN, Wi-Fi, RS232, Machine-to-Machine, OPC UA, Bayern-Hessen protocol and analogue voltage/current outputs |
| GPS | optional IC-ACC-GPSEXT module for geolocation and time synchronisation |
| Included equipment | large zero-air cartridge, PTFE aerosol filter holder with filter, 5 m PFA tubing and 3 m LAN cable |
| Required items | external ACC-PWR-8A power supply; ACC-WIFIUSB module when Wi-Fi connectivity is required |
| Options and accessories | operator tablet, GPS, ICAD cooling, integrated NDIR CO₂ sensor, PTFE filters, regular- or high-capacity zero-air system, maintenance set and IC-SPARE-SET-HONO spare-part set |
| Patents | EP3329251; D502016003239.4; GB3329251; FR3329251; US10473583B2; ZL201680057099.6; DE102015000423; EP325928; D502016014735.3; GB3325928; FR3325928 |
Simultaneous HONO and NO₂ measurement reveals processes that a conventional NOx-only analyser cannot resolve.