AIRYX · ICAD series · NO₂ analyser

ICAD-NO₂-210

A high-precision in-situ analyser for direct nitrogen dioxide measurement. Selective spectral analysis at approximately 430-480 nm eliminates the need for analyte conversion and sample drying. It is suited to air-quality monitoring and NO₂ monitoring around industrial processes.

AIRYX ICAD-NO2-210 analyserICAD-NO₂-210
0.09 ppb2σ detection limit at 60 s averaging
<2 sresponse time at 2 l/min sample flow
<1 mintypical start-up time
<30 Wtypical power consumption at 12 V
Direct NO₂ detection

A characteristic spectrum instead of an indirect reaction

The optical resonator provides an absorption path of up to 2 km. The ICAD algorithm analyses the high-resolution NO₂ spectrum and separates the effects of water, ozone and glyoxal.

Measurement without sample conversion

The ICAD-NO₂-210 uses a single path for direct spectroscopic NO₂ detection. It requires neither an NO converter nor consumable gases.

selective ICAD spectroscopy at approximately 430-480 nm
no sample drying, such as Nafion tubing, required
automatic zero measurements reduce noise and drift
adjustable sample flow from 1 to 2 l/min
insensitive to ordinary vibration and absolute light-source intensity
ICAD analyser measurement principle
Operation and diagnostics

Full control from the panel or via Wi-Fi

The embedded computer performs analysis, data storage and diagnostics. The interface is available from a PC, tablet or phone without installing an additional app.

AIRYX ICAD-NO2-210 user interface

Interface and system health

Graphs and the diagnostics table show gas flow, supply voltage and temperatures of critical components.

AIRYX ICAD analyser connections

System integration

LAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen protocol and optional voltage and current outputs.

Instrument technical data

ICAD-NO₂-210 specification

These parameters apply to the direct NO₂ model. A reduced measurement range can improve precision and the detection limit.

Analytical performance
Measured componentNO₂ - direct ICAD spectroscopy
Measurement range0-5000 ppb; custom configurations available
Time resolutionminimum averaging time 1 s
Delivery time setting5 s as standard; adjustable in the Web UI, with longer settings improving precision through true internal spectral averaging
Detection limit, 2σ0.09 ppb at 60 s
Precision, 1σ0.40 ppb at 1 s; 0.05 ppb at 60 s
Response time, 10-90%below 2 s at 2 l/min; adjustable flow
Zero driftbelow 0.1 ppb/month; automated zero measurements reduce noise and drift
Temperature sensitivitybelow 0.01 ppb/°C
Spectral rangeapproximately 430-480 nm
Effective optical pathup to 2 km
Cross sensitivityno significant spectroscopic cross sensitivity to carbon oxides, formaldehyde, ozone, methane, hydrogen, sulphur compounds, chlorine, ammonia and selected hydrocarbons; glyoxal is resolved spectrally
Sample and calibration
Sample flow1-2 l/min
Sample preparationsample drying is not required
Path-length characterisationhelium every 1-2 years as an option or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version
NO₂ calibrationNO₂ calibration gas is not required; concentration is derived using literature absorption data
Consumable gasesnone
Operating conditions and construction
Operating temperature-10 to +25°C; up to +40°C with optional cooling
Start-up timetypically below 1 min
Power and consumption12 V DC; typically below 30 W
Mechanical stabilityinsensitive to ordinary vibration
Weightbelow 10 kg, depending on configuration
19-inch rack enclosure43.8 × 13.3 cm (3U) × 43.5 cm
IP64 enclosure43.6 × 13.2 × 30.0 cm
Control and communication
Processing unitembedded PC with analysis and measurement software
User interfaceinstrument panel and browser interface accessible through Wi-Fi
Data interfacesLAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen protocol; optional voltage and current outputs
Diagnosticssystem health table, gas-flow, supply-voltage and temperature checks; span adjustment in the interface
Data storageinternal memory for up to approximately two years of data
GPSoptional IC-ACC-GPSEXT module for geolocation and time synchronisation
Delivery, equipment and options
Included equipmentlarge zero-air cartridge, PTFE aerosol filter holder with filter, 5 m PTFE tubing and 3 m LAN cable
Required itemsACC-PWR-8A external power supply; ACC-WIFIUSB module when Wi-Fi communication is required
Options and accessoriesoperator tablet, GPS, ICAD cooling, integrated NDIR CO₂ sensor, PTFE filters, regular- or high-capacity zero-air system, maintenance kit, spare parts and service options
Platform patentsEP3329251; DE 10 2015 000 423; US15/748,923; China ZL201680057099.6
Installation format

For stations and field campaigns

The same measurement model can be supplied in a rack enclosure or a rugged housing for mobile applications.

Rack-mounted ICAD-NO2-210

19-inch rack

For monitoring stations, laboratories and instrument cabinets, with an OLED front display.

ICAD-NO2-210 in an IP64 field enclosure

IP64 field enclosure

A rugged configuration for mobile campaigns and temporary outdoor installations.

Model applications

Where ICAD-NO₂-210 fits

The model supports tasks that require selective NO₂ measurement, high dynamics and low operating effort.

Ambient air qualitystationary monitoring of trace NO₂ concentrations
Industrial monitoringNO₂ control around processes, machinery and installations
Workplace monitoringmeasurements in mines, construction sites and production areas
Scientific researchatmospheric experiments and simulation chambers
Field campaignsmobile measurements, mapping and rapid location comparison