IP64 enclosure
A dust- and weather-resistant construction with connectors suited to operation outside the laboratory.
A mobile Twin Inlet analyser for simultaneous NO₂ measurement from two independent sampling points. Two inlets and two ICAD cells allow direct comparison of both streams without valve switching or a time offset.
ICAD-NO₂-210DT-M
Each sample line has its own inlet and measurement cell. The results are synchronised, allowing two locations or the concentration before and after a test object to be compared directly.
Both paths determine NO₂ directly from its absorption spectrum at approximately 430-480 nm. The instrument does not measure NO or NOx - both channels are dedicated to selective NO₂ measurement.
The UV-resistant polycarbonate enclosure protects the system during mobile campaigns and temporary outdoor installations.

A dust- and weather-resistant construction with connectors suited to operation outside the laboratory.

Both-channel results, time series and diagnostics are available from a PC, tablet or phone without an additional app.
Parameters for the dual-channel Twin Inlet configuration in the mobile M enclosure.
| Measurement arrangement | |
|---|---|
| Measured component | NO₂ on both channels - direct ICAD spectroscopy |
| Sample inlets | 2 independent inlets for parallel sampling from two lines |
| Measurement cells | 2 independent ICAD cells operating simultaneously |
| Measurement range | 0-5000 ppb per channel; custom configurations available |
| Time resolution | minimum averaging time 1 s; typical delivery setting 5 s |
| Detection limit, 2σ | approximately 0.09 ppb at 60 s for a standard Series 210 cell |
| Precision, 1σ | up to 0.40 ppb at 1 s and up to 0.05 ppb at 60 s |
| Response time, 10-90% | below 2 s at 2 l/min; adjustable flow |
| Zero drift | below 0.1 ppb/month; regular automated zero measurements |
| Temperature sensitivity | below 0.01 ppb/°C |
| Spectral range | approximately 430-480 nm |
| Effective optical path | up to 2 km in the resonator cell |
| Cross sensitivity | effects from water, ozone and glyoxal are eliminated or resolved spectrally; no significant cross sensitivity to typical matrix components |
| Sample and calibration | |
| Sample flow | 1-2 l/min, adjustable; independent lines for both sampling points |
| Sample preparation | sample drying is not required; PTFE aerosol filters are used |
| Path-length characterisation | helium every 1-2 years as an option or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version |
| NO₂ calibration | NO₂ calibration gas is not required; concentration is derived using literature absorption data |
| Zero-air system | integrated filter; regular 115 ml or high-capacity 400 ml version available |
| Consumable gases | no gases required for normal operation |
| Enclosure and operating conditions | |
| Enclosure | UV-resistant polycarbonate, IP64, dust- and weather-resistant construction |
| Dimensions | 43.6 × 13.2 × 30.0 cm |
| Weight | below 10 kg, depending on configuration |
| Operating temperature | -10 to +25°C; up to approximately +40°C with optional cooling |
| Start-up time | typically below 1 min |
| Power | 12 V DC; external power supply selected for the configuration |
| Power consumption | typically below 30 W for the Series 210 platform; dependent on optional equipment |
| Mechanical stability | insensitive to ordinary vibration and absolute light-source intensity |
| Control, communication and equipment | |
| Processing unit | embedded PC with analysis and measurement software |
| User interface | browser interface accessible through Wi-Fi on a PC, tablet or phone |
| Data interfaces | LAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen protocol; optional voltage and current outputs |
| Geolocation | optional external GPS with position storage and time synchronisation |
| Diagnostics | gas-flow, supply-voltage and critical-temperature checks; span adjustment in the interface |
| Service equipment | PTFE aerosol filters, maintenance kit, zero-air cartridges and an NO₂-path spare-parts kit |
| Options | operator tablet, GPS, cell cooling, high-capacity zero filter and an integrated CO₂ sensor selected for the application |
The dual-path arrangement removes the timing error caused by sequential point switching and captures rapid concentration changes.
We will select both line lengths, filtration, range, data logging, GPS and equipment for the locations being compared.