AIRYX · ICAD series · NO/NO₂/NOx analyser

ICAD-NOx-210D

A dual-cell analyser for parallel NO and NO₂ measurement and determination of NOx. The first cell measures NO₂ directly by ICAD, while the second detects NO after selective conversion to NO₂. It is suited to air-quality monitoring and dynamic nitrogen-oxide measurements from combustion processes.

AIRYX ICAD-NOx-210D analyserICAD-NOx-210D
2 cellsparallel NO₂ and converted-NO detection
0.09 ppbNO₂-channel 2σ LOD at 60 s averaging
<2 sresponse time at 2 l/min sample flow
<30 Wtypical power consumption at 12 V
Parallel measurement

Direct NO₂ and selectively converted NO

The two measurement cells operate simultaneously. The analyser therefore does not derive NO by subtracting sequential, time-shifted measurements and can resolve rapid changes in sample composition.

How three results are obtained

One path uses ICAD spectroscopy to measure NO₂ at approximately 430-480 nm. In the second, a patented NOx-free ozone generator converts NO to NO₂. The parallel signals provide NO, NO₂ and total NOx concentrations.

two cells for parallel analysis
direct, selective NO₂ detection by ICAD
internal NO conversion using an NOx-free ozone generator
no sample drying required
automated zero measurements reduce noise and drift
ICAD NOx analyser measurement paths
Operation and integration

On-instrument diagnostics and Wi-Fi access

The embedded computer performs analysis, data storage and system health monitoring. The interface is available from a PC, tablet or phone without installing an additional app.

AIRYX ICAD analyser interface

Interface and diagnostics

Graphs and the system-health table show results, gas flow, supply voltage and temperatures of critical components.

AIRYX ICAD analyser connections

System communication

LAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen protocol and optional voltage and current outputs.

Instrument technical data

ICAD-NOx-210D specification

Parameters for the D model measuring NO and NO₂. The stated LOD and precision apply to the direct NO₂ channel.

Analytical performance
Measured componentsNO₂ - direct ICAD spectroscopy; NO - conversion to NO₂; calculated total NOx
Measurement arrangementtwo cells for parallel detection of NO₂ and converted NO
Measurement range0-5000 ppb; custom configurations available
Time resolutionminimum averaging time 1 s
Delivery time setting5 s as standard; adjustable in the Web UI, with longer settings improving precision through true internal spectral averaging
Detection limit, 2σ0.09 ppb at 60 s for the NO₂ channel
Precision, 1σ0.40 ppb at 1 s; 0.05 ppb at 60 s for the NO₂ channel
Response time, 10-90%below 2 s at 2 l/min; adjustable flow
Zero driftbelow 0.1 ppb/month; automated zero measurements reduce noise and drift
Temperature sensitivitybelow 0.01 ppb/°C
Spectral rangeapproximately 430-480 nm
Effective optical pathup to 2 km
Cross sensitivityno significant spectroscopic cross sensitivity to carbon oxides, formaldehyde, ozone, methane, hydrogen, sulphur compounds, chlorine, ammonia and selected hydrocarbons; glyoxal is resolved spectrally
Sample, conversion and calibration
Sample flow1-2 l/min
Sample preparationsample drying is not required
NO conversionpatented internal NOx-free ozone generator; NO is converted to NO₂ in the second path
Path-length characterisationhelium every 1-2 years as an option or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version
NO₂ calibrationNO₂ calibration gas is not required; concentration is derived using literature absorption data
Converter calibrationNO gas above 1000 ppb, approximately every 12 months
Consumable gasesno gases required for normal operation
Operating conditions and construction
Operating temperature-10 to +25°C; up to +40°C with optional cooling
Start-up timetypically below 1 min
Power and consumption12 V DC; typically below 30 W
Mechanical stabilityinsensitive to ordinary vibration
Weightbelow 10 kg, depending on configuration
19-inch rack enclosure43.8 × 13.3 cm (3U) × 43.5 cm
IP64 enclosure, M variant43.6 × 13.2 × 30.0 cm
Control and communication
Processing unitembedded PC with analysis and measurement software
User interfaceinstrument panel and browser interface accessible through Wi-Fi
Data interfacesLAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen protocol; optional voltage and current outputs
Diagnosticssystem-health table, gas-flow, supply-voltage and temperature checks; span adjustment in the interface
Data storageinternal memory for up to approximately two years of data
GPSoptional IC-ACC-GPSEXT module for geolocation and time synchronisation
Delivery, equipment and options
Included equipmentlarge zero-air cartridge, PTFE aerosol filter holder with filter, 5 m PTFE tubing and 3 m LAN cable
Required itemsACC-PWR-8A external power supply; ACC-WIFIUSB module when Wi-Fi communication is required
Options and accessoriesoperator tablet, GPS, ICAD cooling, integrated NDIR CO₂ sensor, PTFE filters, regular- or high-capacity zero-air system, maintenance kit, spare parts and service options
Platform patentsEP3329251; DE 10 2015 000 423; US15/748,923; China ZL201680057099.6
Installation format

19-inch rack or mobile M variant

The standard ICAD-NOx-210D is rack-mounted. Its corresponding M variant uses an IP64 enclosure for field installations.

Rack-mounted ICAD-NOx-210D

19-inch rack

For monitoring stations, laboratories and instrument cabinets, with an OLED front display.

Mobile AIRYX ICAD-NOx-210D-M variant

M variant · IP64

The corresponding configuration for mobile campaigns and temporary outdoor installations.

Model applications

Where ICAD-NOx-210D fits

The D model is intended for fast NO, NO₂ and NOx measurement without an additional CO₂ channel. It also fits mobile campaigns and road-traffic concentration studies. The dedicated Plume Chasing configuration with CO₂ measurement and emission-analysis software is ICAD-NOx-210DE.

Ambient air qualitycontinuous monitoring of trace NO, NO₂ and NOx concentrations
Emissions and combustiondynamic nitrogen-oxide measurement in combustion gases
Industrial monitoringNOx control around processes, machinery and installations
Scientific researchatmospheric experiments, simulation chambers and method validation
Field campaignsmobile measurements and tracking rapid concentration changes
Dedicated Plume Chasing configuration

For dedicated vehicle emission measurements using the Plume Chasing method, the recommended configuration is the ICAD-NOx-210DE, featuring integrated CO₂ measurement and vehicle emission analysis software.

Enquire about ICAD-NOx-210D

We will define the range, enclosure, sample conditions, interfaces and equipment required for the installation.