AIRYX · ICAD series · High Grade NO, NO₂ and NOx

ICAD-NOx-210DL

A high-sensitivity in-situ analyser with two parallel cells: the first measures NO₂ directly, while the second determines NO after internal conversion to NO₂. The DL version combines a complete NO, NO₂ and NOx balance with the sensitivity required for background studies and reference measurements.

AIRYX ICAD-NOx-210DL analyserICAD-NOx-210DL
0.03 ppb2σ detection limit at 60 s
0.02 ppb1σ precision at 60 s
<3 sresponse time at 2 l/min
up to 4 kmeffective absorption path
Two measurement paths

Direct NO₂ and parallel NO determination

Patented ICAD spectroscopy identifies NO₂ around 430–480 nm. The second cell receives sample after NO has been converted to NO₂ in an internal NOx-free ozone generator, allowing the instrument to report NO, NO₂ and total NOx simultaneously.

Parallel measurement without conventional converter artefacts

two optical resonator cells operating in parallel
direct NO₂ detection by ICAD spectroscopy
internal NO → NO₂ conversion without adding NOx
no sample drying required
optional NDIR CO₂ sensor in the sample path
integrated zero measurements to limit drift
Dual measurement-path and NO-converter diagram for the ICAD-NOx-210DL
DL version — High Grade

Sensitivity for very low NOx levels

Extended cells with High Grade mirrors create an optical path of up to 4 km. Active Peltier cooling stabilises LED temperature to better than 0.1°C, improving accuracy and reducing long-term drift.

ICAD-NOx-210DL NO2 precision versus averaging time

Precision improves with averaging

1σ precision is 0.12 ppb at 1 s and reaches 0.02 ppb at 60 s. The standard delivery setting is 30 s and can be changed in the Web UI.

AIRYX ICAD-NOx-210DL user interface on a tablet

Control without additional software

The browser interface displays NO, NO₂ and NOx time series, settings and the System Health table. It can be accessed from a computer, tablet or phone.

Stability and diagnostics

Designed for extended measurement campaigns

The result is independent of absolute light-source intensity, while the optical system is insensitive to typical vibration. On-board diagnostics monitor parameters including flow, supply voltage and temperature.

◎

Automatic zero

Regular reference measurements reduce noise and maintain zero drift below 0.1 ppb/month.

↯

Fast start and response

Typical start-up is under one minute, with a 10–90% response below 3 s at 2 l/min.

⌁

Multi-channel integration

LAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen and analogue voltage/current outputs.

Instrument technical data

ICAD-NOx-210DL specification

Parameters refer to the dual-path High Grade version in a 19″ rack enclosure. Range and configuration can be adapted to the application, and a reduced range can improve precision and the detection limit.

Analytical performance
Measured speciesNO₂ — direct ICAD measurement; NO — after conversion to NO₂; NOx — total NO + NO₂ result
Measurement range0–2000 ppb; custom configurations and ranges are available
Detection limit, 2σ0.03 ppb at 60 s averaging
Precision, 1σ0.12 ppb at 1 s; 0.02 ppb at 60 s
Time resolution1 s minimum averaging; 30 s standard delivery setting
Response time, 10–90%less than 3 s at a 2 l/min sample flow; flow is adjustable
Zero driftless than 0.1 ppb/month; automatic zero measurements further reduce noise and drift
Temperature sensitivityless than 0.01 ppb/°C
Spectral rangeapproximately 430–480 nm
Absorption pathup to 4 km in the optical resonators
Measurement cellstwo parallel enhanced-sensitivity cells, almost twice as long as the standard version and fitted with High Grade mirrors
NO₂ detectiondirect ICAD spectroscopy using the characteristic NO₂ spectrum
NO and NOx detectionNO is converted to NO₂ in a patented NOx-free ozone generator; the second path determines NO and the instrument calculates NOx
Light-source stabilisationactive Peltier LED cooling; temperature stability better than 0.1°C
Other detectable gasesglyoxal; CO₂ with an optional NDIR sensor
Cross-sensitivityno significant spectroscopic cross-sensitivity to, among others, carbon oxides, formaldehyde, ozone, methane, hydrogen, sulphide, SO₂, chlorine, ClO₂, HCN, HCl, phosphine, ammonia, acetylene, nitromethane, ethylene, ethanol, methyl mercaptan or ethyl mercaptan
Sample, zero and calibration
Sample flow1–2 l/min
Sample preparationsample drying is not required; a PTFE aerosol filter is used at the inlet
NO₂ calibrationNO₂ calibration gas is not required; concentration is quantified from literature absorption data
NO-converter checkfor concentrations above 1000 ppb, a periodic check with NO gas is required, approximately every 12 months
Optical-path characterisationoptional helium every 1–2 years or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version
Zero-air systemlarge zero-air cartridge included; regular- and high-capacity systems are available
Consumable gasesnone required during normal operation; NO gas is associated with periodic converter checking under the conditions above
Construction and operating conditions
Enclosure19″ rack, 3U high, with a front-panel OLED display
Dimensions43.8 × 13.3 × 66.5 cm (W × H × D)
Weightless than 12 kg, depending on configuration
Operating temperature-10 to +40°C
Mechanical stabilityinsensitive to typical vibration
Power12 V DC; external ACC-PWR-8A power supply required
Power consumptiontypically less than 40 W at 12 V
Start-up timetypically less than 1 min
Mobile usethe rack instrument can be used in suitably mounted mobile installations; it is not the IP64 M field version
Control, communication and equipment
Processing unitembedded industrial computer with data-analysis and measurement software
User interfaceWeb UI with time series, span-calibration adjustment, System Health table and on-board diagnostics; accessible from a PC, tablet or phone
Data interfacesLAN, Wi-Fi, RS232, Machine-to-Machine, OPC UA, Bayern-Hessen protocol and analogue voltage/current outputs
GPSoptional IC-ACC-GPSEXT module for geolocation and time synchronisation
Included equipmentlarge zero-air cartridge, PTFE aerosol filter holder with filter, 5 m PFA tubing and 3 m LAN cable
Required itemsexternal ACC-PWR-8A power supply; ACC-WIFIUSB module when wireless connectivity is required
Options and accessoriesoperator tablet, GPS, ICAD cooling, integrated NDIR CO₂ sensor, PTFE filters, regular- or high-capacity zero-air system, maintenance set and IC-SPARE-SET-NOX spare-part set
Family variantsICAD-NO₂-210L — NO₂ only; ICAD-NOx-210DL — NO, NO₂ and NOx
PatentsEP3329251; D502016003239.4; GB3329251; FR3329251; US10473583B2; ZL201680057099.6; DE102015000423; EP325928; D502016014735.3; GB3325928; FR3325928
210DL applications

Complete NOx analysis at background levels

The DL version is intended for measurements that require NO and NO₂ to be resolved while retaining the precision associated with High Grade analysers.

Atmospheric and clean-background researchppt-level NO, NO₂ and NOx measurements and long observation campaigns
Air-quality stationsurban, regional and roadside background monitoring with ample sensitivity reserve
Laboratories and test bencheschamber studies, method validation, combustion-source tests and reference measurements
Emission controlprecise separation of NO and NO₂ in engine, combustion-process and industrial-source studies