Precision improves with averaging
1σ precision is 0.12 ppb at 1 s and reaches 0.02 ppb at 60 s. The standard delivery setting is 30 s and can be changed in the Web UI.
A high-sensitivity in-situ analyser with two parallel cells: the first measures NO₂ directly, while the second determines NO after internal conversion to NO₂. The DL version combines a complete NO, NO₂ and NOx balance with the sensitivity required for background studies and reference measurements.
ICAD-NOx-210DL
Patented ICAD spectroscopy identifies NO₂ around 430–480 nm. The second cell receives sample after NO has been converted to NO₂ in an internal NOx-free ozone generator, allowing the instrument to report NO, NO₂ and total NOx simultaneously.
Extended cells with High Grade mirrors create an optical path of up to 4 km. Active Peltier cooling stabilises LED temperature to better than 0.1°C, improving accuracy and reducing long-term drift.
1σ precision is 0.12 ppb at 1 s and reaches 0.02 ppb at 60 s. The standard delivery setting is 30 s and can be changed in the Web UI.
The browser interface displays NO, NO₂ and NOx time series, settings and the System Health table. It can be accessed from a computer, tablet or phone.
The result is independent of absolute light-source intensity, while the optical system is insensitive to typical vibration. On-board diagnostics monitor parameters including flow, supply voltage and temperature.
Regular reference measurements reduce noise and maintain zero drift below 0.1 ppb/month.
Typical start-up is under one minute, with a 10–90% response below 3 s at 2 l/min.
LAN, Wi-Fi, RS232, M2M, OPC UA, Bayern-Hessen and analogue voltage/current outputs.
Parameters refer to the dual-path High Grade version in a 19″ rack enclosure. Range and configuration can be adapted to the application, and a reduced range can improve precision and the detection limit.
| Analytical performance | |
|---|---|
| Measured species | NO₂ — direct ICAD measurement; NO — after conversion to NO₂; NOx — total NO + NO₂ result |
| Measurement range | 0–2000 ppb; custom configurations and ranges are available |
| Detection limit, 2σ | 0.03 ppb at 60 s averaging |
| Precision, 1σ | 0.12 ppb at 1 s; 0.02 ppb at 60 s |
| Time resolution | 1 s minimum averaging; 30 s standard delivery setting |
| Response time, 10–90% | less than 3 s at a 2 l/min sample flow; flow is adjustable |
| Zero drift | less than 0.1 ppb/month; automatic zero measurements further reduce noise and drift |
| Temperature sensitivity | less than 0.01 ppb/°C |
| Spectral range | approximately 430–480 nm |
| Absorption path | up to 4 km in the optical resonators |
| Measurement cells | two parallel enhanced-sensitivity cells, almost twice as long as the standard version and fitted with High Grade mirrors |
| NO₂ detection | direct ICAD spectroscopy using the characteristic NO₂ spectrum |
| NO and NOx detection | NO is converted to NO₂ in a patented NOx-free ozone generator; the second path determines NO and the instrument calculates NOx |
| Light-source stabilisation | active Peltier LED cooling; temperature stability better than 0.1°C |
| Other detectable gases | glyoxal; CO₂ with an optional NDIR sensor |
| Cross-sensitivity | no significant spectroscopic cross-sensitivity to, among others, carbon oxides, formaldehyde, ozone, methane, hydrogen, sulphide, SO₂, chlorine, ClO₂, HCN, HCl, phosphine, ammonia, acetylene, nitromethane, ethylene, ethanol, methyl mercaptan or ethyl mercaptan |
| Sample, zero and calibration | |
| Sample flow | 1–2 l/min |
| Sample preparation | sample drying is not required; a PTFE aerosol filter is used at the inlet |
| NO₂ calibration | NO₂ calibration gas is not required; concentration is quantified from literature absorption data |
| NO-converter check | for concentrations above 1000 ppb, a periodic check with NO gas is required, approximately every 12 months |
| Optical-path characterisation | optional helium every 1–2 years or an NO₂ reference measurement; gas-free ICOM method subject to confirmation of availability in the current software version |
| Zero-air system | large zero-air cartridge included; regular- and high-capacity systems are available |
| Consumable gases | none required during normal operation; NO gas is associated with periodic converter checking under the conditions above |
| Construction and operating conditions | |
| Enclosure | 19″ rack, 3U high, with a front-panel OLED display |
| Dimensions | 43.8 × 13.3 × 66.5 cm (W × H × D) |
| Weight | less than 12 kg, depending on configuration |
| Operating temperature | -10 to +40°C |
| Mechanical stability | insensitive to typical vibration |
| Power | 12 V DC; external ACC-PWR-8A power supply required |
| Power consumption | typically less than 40 W at 12 V |
| Start-up time | typically less than 1 min |
| Mobile use | the rack instrument can be used in suitably mounted mobile installations; it is not the IP64 M field version |
| Control, communication and equipment | |
| Processing unit | embedded industrial computer with data-analysis and measurement software |
| User interface | Web UI with time series, span-calibration adjustment, System Health table and on-board diagnostics; accessible from a PC, tablet or phone |
| Data interfaces | LAN, Wi-Fi, RS232, Machine-to-Machine, OPC UA, Bayern-Hessen protocol and analogue voltage/current outputs |
| GPS | optional IC-ACC-GPSEXT module for geolocation and time synchronisation |
| Included equipment | large zero-air cartridge, PTFE aerosol filter holder with filter, 5 m PFA tubing and 3 m LAN cable |
| Required items | external ACC-PWR-8A power supply; ACC-WIFIUSB module when wireless connectivity is required |
| Options and accessories | operator tablet, GPS, ICAD cooling, integrated NDIR CO₂ sensor, PTFE filters, regular- or high-capacity zero-air system, maintenance set and IC-SPARE-SET-NOX spare-part set |
| Family variants | ICAD-NO₂-210L — NO₂ only; ICAD-NOx-210DL — NO, NO₂ and NOx |
| Patents | EP3329251; D502016003239.4; GB3329251; FR3329251; US10473583B2; ZL201680057099.6; DE102015000423; EP325928; D502016014735.3; GB3325928; FR3325928 |
The DL version is intended for measurements that require NO and NO₂ to be resolved while retaining the precision associated with High Grade analysers.
We will select the range, averaging time, converter-check procedure, communication, CO₂ sensor and other equipment for the planned NOx measurement.