High-power LEDs
The standard light source uses three LEDs centred at 275, 310 and 370 nm.
- 270–380 nm spectrometer
- typically 0.45 nm FWHM
- time resolution up to 3 s
An active, modular Long-Path DOAS system for continuous path-averaged gas measurements. Its separate telescope, stabilised analyser and retroreflector let the optical configuration match each installation.
OPEN PATH HIGH GRADEThe analyser is installed indoors or in a sheltered position and connected to the telescope by cables and optical fibres. Separating the components allows the optics to be positioned correctly in the field.

Select the source to match the required species and spectral range.
The standard light source uses three LEDs centred at 275, 310 and 370 nm.
A broadband Xe lamp / LDLS extends the spectral range and maximum measurement-path length.
The final species list and detection limits depend on source, path length, atmospheric conditions and averaging time.
Values describe the current Open Path family; confirm the final configuration for each project.
| Measurement | |
|---|---|
| Method | active Long-Path DOAS absorption spectroscopy |
| One-way path length | 100–3,000 m; 500–1,500 m recommended |
| Time resolution | up to 3 s; adjustable averaging improves sensitivity |
| Calibration | no calibration gases; quantification uses literature absorption data |
| Lamp reference | automatic, using a reference plate in front of the fibre bundle |
| Telescope | |
| Optics | 800 mm focal length; 200 mm mirror; approximately 0.05° field of view |
| Path alignment | camera, fibre-bundle motors and automatic optimisation |
| Fibre configuration | 6 sending mono fibres and 1 receiving mono fibre; UV fused-silica multi-fibre bundle |
| Enclosure | IP64; water- and snow-resistant design |
| Weight and dimensions | approximately 20 kg; 110 × 40 × 36 cm |
| Operating temperature | −20 to +40°C |
| Mounting | aluminium rail system or tripod |
| Analyser and data | |
| Spectrometer | 270–380 nm; typically 0.45 nm FWHM; ultra-low stray-light configuration |
| Detector quantum efficiency | approximately 60%, back-UV-thinned detector |
| Noise | 10⁻⁴ at 1,000 scans, approximately 60 s integration |
| Light sources | LEDs at 275, 310 and 370 nm; optional broadband Xe lamp / LDLS |
| Stabilisation | temperature-stabilised spectrometer and light sources |
| Weight and dimensions | approximately 14 kg; 40 × 44 × 13 cm |
| Software | included; configurable routine, averaging and multiple paths |
| Data analysis | package for standard trace gases |
| Communication | USB 2.0; standard operation with an external measurement computer, not included in the standard delivery |
| Sensors | temperature, pressure and humidity |
| Power | 12 V; below 70 W |
| Automation and reflector | |
| Optional drives | azimuth −130° to +130°; elevation 0–30°; automatic multi-path measurements |
| Retroreflector | twenty 1″ fused-silica corner cubes in an IP64 enclosure; array quantity depends on the application and path length |
| Species | Typical LOD at 1,000 m and 10 s |
|---|---|
| NO₂ | 1.5 ppb |
| SO₂ | 0.3 ppb |
| O₃ | 15 ppb |
| HCHO | 4.5 ppb |
| HONO | 0.8 ppb |
High Grade is a configured measurement system rather than a single compact enclosure.
A configuration with a computer embedded in the analyser unit is available as an option.
We will select the light source, species, path length and number, reflectors, drives and mounting method.