AIRYX · Open Path Compact · UV-DOAS

AIRYX OPC-UV3-DL-301

Open Path Ammonia Analyzer

A multi-component Open Path UV-DOAS system for contact-free trace-gas measurement. It analyses NH₃, NO, SO₂, O₃, H₂S, CS₂ and selected VOCs and aromatic compounds in parallel, without sample extraction. Results represent the average concentration across the full optical path, and all measured components are recorded in the output file.

AIRYX OPC-UV3-DL-301 Open Path ammonia analyserOPC-UV3-DL-301
✓
Multi-component UV-DOASparallel analysis of all components in the current configuration
✓
NH₃ up to 100 ppmreference range 0–100 000 ppb
✓
H₂S up to 1.4 ppmreference range 0–1400 ppb
✓
Remote communicationLAN / Wi-Fi / local data storage
Measurement principle

A UV beam instead of sample extraction

Light from a deuterium lamp travels to a retroreflector and back. The DOAS algorithm separates the characteristic gas absorption from scattering, aerosols and intensity variations.

A result representative of the full path

Concentration is averaged across the complete atmospheric path. No probe, pump, filter or heated sampling line is required.

Multi-component measurement. Components are identified by their characteristic absorption spectra and analysed in parallel. With no sample extraction, there are no losses in sampling lines or filters. No calibration gases are required.

Automatic referencing prevents drift, while the method remains robust against lamp ageing, rain, scattering and window losses.

AIRYX Open Path UV-DOAS measurement principle
Data quality

Detectable compounds

Each OPC system analyses all components specified in the current configuration and data sheet, and results for every measured component are recorded in the output file.

ComponentLOD [ppb]Range [ppb]
Inorganic / core gases
NH₃10–100 000
NO20–5000
SO₂50–2000
O₃150–1000
H₂S2000–1400
CS₂0.20–35
VOCs / aromatic and organic compounds
Toluene200–6000
Styrene50–170
Benzene150–26 000
Chlorobenzene120–4400
m-Xylene200–9000
p-Xylene80–3700
m-Cresol40–9000
p-Cresol10–900
o-Cresol60–1800
Benzaldehyde30–1600
Ethylamine100–3200
Methylfurane80–350
Reference conditions in v2.4: detection limits and ranges depend on the optical-path configuration. LOD values refer to a 100 m distance between telescope and reflector and depend on visibility; they are not a universal guarantee for every installation. Ranges are specified for an optical density of 1 at 25 m. Other ranges are available on request.
Multiple paths

One analyser, several measurement directions

The motorised telescope routes the beam between retroreflectors. The camera and searchlight support path identification, automatic and remote alignment, and position correction without manual field aiming.

Technical data

Complete OPC-UV3-DL-301 specifications

Parameters follow the current technical data sheet supplied for the instrument.

Measurement performance
Detectable gasesNH₃, NO, SO₂, O₃, H₂S, CS₂ and the VOCs and aromatic compounds listed in the component table
Detection limits and ranges (v2.4)18 components; full values are listed in the ‘Measurable components’ section above.
Time resolutionup to 1 minute
Reference conditionsLOD depends on the path configuration and visibility; values refer to a 100 m distance between telescope and reflector. Ranges: optical density of 1 at 25 m; other ranges available on request.
Optics
Light sourcedeuterium lamp; lifetime over 3 months, typically 6 months
Telescopemirror telescope; 300 mm focal length
Spectrometer190–310 nm; resolution <0.5 nm FWHM; temperature stabilised
Distance to reflector20–150 m
Construction and maintenance
WeatherproofingIP64; internal heating; dew-point monitoring; window wiper; suitable for outdoor installation
Installationstationary or mobile on an optional tripod; multiple light paths supported
Routine maintenancelamp replacement 2–3 times per year; desiccant replacement twice per year; site-dependent cleaning of window, retroreflector and heat-sink filter
Consumablesno other consumables besides lamp and desiccant
Motorisationazimuth 0–360°; elevation −40–75°
Temperature controltwo-stage heating and cooling of spectrometer and housing
Smart setupintegrated camera and searchlight; automatic and remote light-path alignment
Additional sensorsambient temperature and pressure; internal dew point
Control and communication
Control unitembedded PC with Windows 11 IoT; measurement control and analysis run on the instrument’s internal PC
Data interfacesinternal data storage; LAN; Wi‑Fi; M2M communication and UIDEP protocol support
Result recordingResults for all analysed components are recorded in an output file in the instrument’s internal storage.
Power supply and consumption24 V; 80 W max.
Long-term operation

Field-campaign evidence

AIRYX materials show a one-year campaign at Neuglobsow with over 90% data coverage and comparisons of NH₃ measurements against established methods.

AIRYX Open Path one-year measurement campaign at Neuglobsow
Two paths of 79 and 130 m; an example of the observed NH₃ diurnal pattern.
AIRYX Open Path comparison against filterpack and denuder methods
Correlation with filterpack: R² = 0.85; correlation with denuder: R² = 0.82.
Applications

From farms to facility fence lines

Multi-component Open Path measurement supports air-quality monitoring, agricultural and industrial emissions monitoring, process control and analysis of the spatial distribution of trace gases. H₂S can be included in industrial and facility-boundary (fence-line) monitoring.

AgricultureNH₃ and diffuse sources in agriculture
Industrial emissionsNO, SO₂, H₂S, CS₂ and VOCs from industrial sources
Fence-line monitoringincluding H₂S, NH₃ and VOCs at facility boundaries
Air qualityO₃, NO, SO₂, NH₃ and VOCs in ambient air
VOCs and aromatic compoundsselective measurement of organic compounds including benzene, toluene, styrene and xylenes
Process controldynamic multi-component Open Path measurement
Spatial analysistrace-gas distribution along one or several paths