AIRYX · SkySpec · stationary modular MAX-DOAS

SkySpec 1D

A modular system combining a separate outdoor telescope with a temperature-stabilised spectrometer for automatic scattered-sunlight observations. Motorised elevation, fixed azimuth and up to two UV/VIS channels form a platform for long-term atmospheric monitoring.

AIRYX SkySpec 1D v.260 telescopeSKYSPEC 1D
−10° to 190°programmable elevation range
< 0.2°automatic elevation correction accuracy, 1σ
< 2 mincomplete-system start-up time
IP64telescope and SpecBox for demanding conditions
Measurement principle

One axis, automatic angular sequences

SkySpec 1D records spectra of scattered skylight at programmed elevation angles. DOAS analysis determines gas contents integrated along the optical path, vertical columns and near-surface concentrations.

Telescope outdoors, analysis indoors

The outdoor telescope is connected by optical fibre to the SpecBox and measurement computer installed indoors or in a protected location.

motorised elevation axis and fixed azimuth direction
inclinometer and real-time position correction
narrow field of view below 0.3° × 1°
no external moving telescope parts
compliant with VDI 4212 requirements for passive DOAS
v.260: faster elevation drive and a shorter, more robust tube
Complete SkySpec 1D system with telescope, SpecBox and computer
System architecture

1D telescope and configurable SpecBox

The separated architecture supports either one broadband spectrometer or two optimised UV and VIS channels, together with a suitable optical-fibre design and length.

AIRYX SkySpec 1D architecture

Stable optics for long campaigns

The fibre cross-section is transformed from circular to slit-shaped, improving throughput and providing uniform spectrometer illumination.

active spectrometer temperature stabilisation
low stray light and high spectral sampling
mirror-free optics limiting polarisation errors
telescope heating activated below 5°C
optional Hg lamp and shutter for calibration checks
Measured components

Trace gases and aerosol properties

The available components depend on the spectrometer, detector, filter and Evaluation-package configuration.

NO₂nitrogen dioxide
O₃ozone
SO₂sulphur dioxide
HCHOformaldehyde
H₂Owater vapour
HONOnitrous acid
BrObromine monoxide
IOiodine monoxide
CHOCHOglyoxal
O₄oxygen collision complex
AODaerosol optical depth
Spectrometer configurations

One channel or a complete UV/VIS system

SkySpec 1D can use one 300–460 nm channel or two spectrometers optimised for UV and VIS.

SKY-SBOX-1S-RGD / HGD

One spectrometer

A compact configuration typically covering 300–460 nm.

  • RGD CMOS or HGD back-thinned CCD
  • resolution below 0.7 nm FWHM
  • SpecBox weight approximately 8 kg
SKY-SBOX-2S-MGD / RGD / HGD

Two UV/VIS spectrometers

Two stabilised channels for broader coverage and optimised sensitivity.

  • UV: 300–408 nm, below 0.5 nm FWHM
  • VIS: 408–553 nm, below 0.6 nm FWHM
  • mixed HGD UV + RGD VIS option
Technical data

SkySpec 1D v.260 specifications

System, telescope, spectrometer and optical-connection parameters consolidated from AIRYX product sheets and the v1.23 overview.

Spectrometers and measurement
Methodpassive UV/VIS DOAS remote sensing, including automatic MAX-DOAS sequences
UV-I configuration300–408 nm; resolution below 0.5 nm FWHM; BG3 filter
VIS configuration408–553 nm; resolution below 0.6 nm FWHM; BG40 filter
UV-II configuration300–460 nm; resolution below 0.7 nm FWHM; BG3 filter; replaces UV-I
Number of spectrometersup to two; single- and dual-channel configurations available
Noisebelow 3 × 10−4 at 103 scans, approximately 60 s integration
Spectral samplingmore than 6 points across the FWHM of the slit function
Quantum efficiencyUV above 50% with a back-thinned detector; VIS up to 80%
Spectrometer temperature20°C, adjustable; stability better than ±0.03°C
Wavelength calibrationfactory calibration; typical shift below 0.01 nm; optional integrated or handheld Hg lamp
Characterisationwavelength calibration, offset, dark-current spectra and detector non-linearity function
Telescope and geometry
Driveone motorised elevation axis; azimuth set by the mounting orientation
Elevation range−10° to 190°; setting resolution 0.1°
Elevation accuracybelow 0.2° (1σ), automatic inclinometer correction
FOV, FWHMbelow 0.3° × 1°, vertical × horizontal
Telescope spectral range200 nm-2 µm with fused-silica optical components; custom configurations available
Telescope optics75 mm internal focal length; external focus at infinity at 400 nm; 22.5 mm clear aperture; f/4
Heatingautomatic activation below 5°C
Cameraone or two optional 120° × 90° cameras; two cameras cover 0–180° elevation
Operating conditions and construction
Operating temperatureSpecBox −10 to +40°C; telescope −30 to +50°C
EnclosuresIP64; resistant to UV, dust, water and demanding outdoor conditions
Sensorstemperature ±1°C; pressure ±0.5%; relative humidity ±3%
SpecBox weightapproximately 8 kg
Telescope weightapproximately 4 kg
Complete-system weightapproximately 17 kg including laptop, fibres and cables
SpecBox dimensions40 × 30 × 13.2 cm
Telescope dimensionshousing 20 × 20 × 13.2 cm; tube length approximately 12.5 cm
Connections, control and power
Optical fibre10 m as standard in the current overview; 15 or 20 m options and extensions; different cross-sections and connectors
CommunicationSpecBox USB 2.0; telescope RS232 with a USB adapter for stand-alone operation
SoftwareSKY-SOFT-STD for control and acquisition; optional Evaluation package for dSCD, VCD and NSC
System start-upbelow 2 min
Power12 V DC; complete-system consumption typically below 30 W, maximum 100 W
Stand-alone telescoperequires ACC-PWR-8A, SKY-SOFT-TELE and an RS232-to-USB adapter; start-up below 10 s; typically below 2 W, maximum 20 W with heating
Equipment and configuration
Required componentsSKY-TELE-1D telescope, selected SKY-SBOX-* SpecBox, SKY-SOFT-STD, measurement computer and ACC-PWR-8A power supply
Field calibrationoptional shutter or shutter with integrated mercury Hg lamp; handheld lamp also available
Mountingtripod, wall bracket, mast or rail; tripod 0.91–1.69 m, weight 3.4 kg
Supportspare-parts and maintenance kits, installation, configuration and online support

SkySpec 1D is not intended for automatic sun tracking or 2D imaging. AIRYX provides the SkySpec 2D telescope for these tasks.

System equipment

SpecBox, optics and field calibration

The modular architecture allows the system to be built around the required spectral range, installation site and measurement schedule.

AIRYX SpecBox for SkySpec 1D

Single- or dual-channel SpecBox

A temperature-stabilised enclosure with one or two spectrometers and USB communication.

  • UV, VIS or UV-II
  • RGD, HGD or Mixed Grade detector configuration
  • factory-matched optical fibre
Rear section of the AIRYX SkySpec 1D telescope housing

Mounting and accessories

The system can be installed on a tripod, wall, mast or rail.

  • optional field-of-view cameras
  • Hg lamp and automatic shutter
  • transport cases, spare cables and service kits
Applications

Long-term atmospheric monitoring

SkySpec 1D is recommended for stationary measurements and long campaigns requiring stable MAX-DOAS geometry.

Air qualityregional and urban observations of trace gases and aerosols
Vertical profilesMAX-DOAS elevation sequences and external profile inversion
Satellite validationrepresentative ground-based data for Earth-observation products
Point sourcesremote observation of plumes from industrial plants, power stations and volcanoes
Moving platformselevation correction during ship- and vehicle-based measurements
Special researchsurface reflection, plant fluorescence and custom optical systems
Solution selection

Let us configure SkySpec 1D

We will select the telescope, number of spectrometers, detectors, spectral ranges, optical fibre, computer and software for the planned campaign.

AIRYXMES-EKOinstrument and measurement-system selectionmeseko@meseko.pl