AIRYX · SkySpec · mobile passive remote sensing

SkySpec Compact

A self-contained telescope-spectrometer system with an embedded computer for UV/VIS DOAS remote sensing of trace gases and aerosols. Its compact design, automatic elevation correction and 12 V supply support mobile measurements and operation at sites with limited infrastructure.

AIRYX SkySpec Compact v.260 systemSKYSPEC COMPACT
< 0.7 nmstandard FWHM spectral resolution
< 0.2°automatic elevation correction accuracy, 1σ
< 3 minsystem start-up time
< 30 Wtypical power consumption at 12 V
Passive remote sensing

Scattered sunlight instead of sample extraction

SkySpec Compact observes scattered sunlight at programmable elevation angles. Characteristic absorption structures allow DOAS retrieval of light-path gas abundances and aerosol parameters.

Automatic MAX-DOAS sequences

A motor rotates the prism inside the quartz tube, avoiding external moving parts. The integrated inclination sensor corrects the viewing elevation in real time.

motorised elevation scanning from -10° to 190°
fixed azimuth and narrow field of view below 0.3° × 1°
homogeneous slit illumination through an optical fibre
prism arrangement that limits polarisation-induced bias
automatic dark spectra and spectrometer stability control
SkySpec Compact scattered-sunlight observation principle
All-in-one system

Telescope, spectrometer and computer in one enclosure

Compact does not require a separate SpecBox. Control, acquisition, temperature stabilisation and data logging are handled by the embedded computer.

Internal architecture of the AIRYX SkySpec Compact

Stable field optics

The quartz-glass tube protects the optics and is easy to clean. Heating activated below 5°C limits condensation, snow and ice deposition and prevents mechanical components from freezing.

actively temperature-stabilised spectrometer
low-straylight optical design
embedded computer, GPS, LAN and optional Wi-Fi
temperature, pressure and humidity sensors
UV-resistant IP64 polycarbonate enclosure
Measured components

Multiple results from one UV/VIS spectrum

The standard 300-460 nm range can be adapted to the application. The final component set and sensitivity depend on spectral range, detector grade, viewing geometry and evaluation configuration.

NO₂nitrogen dioxide
HCHOformaldehyde
SO₂sulphur dioxide
O₄oxygen collision complex
O₃ozone
H₂Owater vapour
HONOnitrous acid
CHOCHOglyoxal
BrObromine monoxide
IOiodine monoxide
AODaerosol optical depth

O₄ is an oxygen collision complex used, among other purposes, to analyse aerosol properties and the optical path. It should not be interpreted as an independent trace gas.

Detector selection

Compact RGD or Compact HGD

Both versions share the same mechanical design and system functions. They differ in detector grade and UV sensitivity.

SKY-COMPACT-RGD

Regular Grade Detector

A cost-efficient CMOS configuration intended primarily for components with strong absorption in the visible range.

  • recommended for NO₂, H₂O, IO and CHOCHO
  • standard 300-460 nm range
  • spectral resolution below 0.7 nm
SKY-COMPACT-HGD

High Grade Detector

The back-thinned CCD provides more than twice the photon yield of the regular CMOS and maximises UV sensitivity.

  • recommended for O₃, SO₂, HONO, HCHO and BrO
  • UV quantum efficiency above 50%
  • BG3 filter for straylight reduction
Complete instrument data

SkySpec Compact v.260 specifications

Values refer to the standard configuration. Spectral range, filter and detector can be customised within the constraints required for optimum spectroscopic measurements.

Spectrometer and spectral performance
Methodpassive UV/VIS DOAS remote sensing of trace gases and aerosols, including MAX-DOAS sequences
Wavelength range300-460 nm as standard; custom ranges available
Spectral resolutionbelow 0.7 nm FWHM as standard
Optical filterSchott BG3 or BG40 as standard, depending on configuration
Noisebelow 3 × 10-4 at 103 scans, approximately 60 s integration time
Spectral samplingmore than 6 points over the slit-function FWHM
Quantum efficiencyabove 50% in the UV for the HGD back-thinned CCD detector
Spectrometer temperature20°C, adjustable
Temperature stabilitybetter than ±0.03°C
Wavelength calibrationhighly stable in-house calibration, typical shifts below 0.01 nm; manual check or recalibration with a mercury Hg lamp
Spectrometer correctionsoffset and dark-current spectra, wavelength calibration and detector non-linearity function
Telescope and viewing geometry
Drivemotorised viewing-elevation axis, fixed azimuth
Elevation range-10° to 190°
Elevation correctionautomatic real-time inclination-sensor correction; accuracy below 0.2° (1σ)
Field of view FWHMbelow 0.3° × 1°, vertical × horizontal
Lens focal length75 mm
Lens clear aperture22.5 mm
Telescope heatingautomatic activation below 5°C
Cameraoptional; maximum one integrated wide-FOV camera; standard field of view approximately 130° × 90°, customisable on request
Operating conditions, sensors and construction
Operating temperature-10 to +40°C; direct sunlight can raise the enclosure above this range, while higher-temperature operation is possible after adjusting the spectrometer setpoint
Mechanical stabilityIP64 enclosure, sun roof, robust design and simple mounting
Enclosure materialUV-resistant, dust- and weatherproof polycarbonate with outdoor connectors
Internal temperaturestelescope, spectrometer and electronics temperature measurement with 1°C accuracy
Pressureambient-pressure sensor with 0.5% accuracy
Humiditysensor inside the instrument with ±3% relative-humidity accuracy
GPSincluded
Weightapproximately 7 kg
Enclosure dimensions30 × 20 × 13.2 cm, box only
Quartz tubeapproximately 12.5 cm long and 8 cm in diameter
Control, communication and power
Computerembedded, preconfigured computer with Windows 10 Professional
Measurement softwarecustomisable angle and time-resolution routine; measurements start automatically after power-up
CommunicationLAN; Wi-Fi through a dedicated external adapter
Data cable2 m LAN cable with IP68 connector included with the instrument
Start-up timebelow 3 min
Power supply12 V DC; platform supply range 9-15 V
Power consumptiontypically below 30 W; maximum 100 W
Mobile poweroptional 50 Ah / 13.6 V LiPo battery in a Peli case or vehicle electrical supply
Configuration and equipment
Detector variantsRegular Grade CMOS (RGD) or High Grade back-thinned CCD (HGD)
Required configurationSKY-SOFT-STD control and acquisition software, dedicated ACC-WIFIUSB Wi-Fi module and ACC-PWR-8A external power supply
SKY-ACC-TRIPODheight 0.91-1.69 m; leg spacing 0.80-1.35 m; weight 3.4 kg; transport length 1.06 m
Mounting optionsoutdoor tripod, rail and mast adapters, and custom mounting components
Optical optionshandheld mercury Hg wavelength lamp, wide-FOV camera, custom spectrometer range and filters
Evaluation softwareEvaluation package for automatic near-real-time and post-processing analysis, visualisation and data export
Maintenance supportspare-parts set, maintenance set, tablet, installation and online support
Field configuration

A complete kit for mobile operation

Low weight and power consumption support a self-contained station with tripod and battery. LAN or Wi-Fi access enables operation from a computer, tablet or phone.

SkySpec Compact mounted on an outdoor tripod

Tripod and mounting

The outdoor tripod places the instrument at the required height. Rail and mast adapters are also available.

  • no external moving parts
  • easy quartz-tube cleaning
  • fast start after setting the viewing direction
SkySpec Compact field kit with tripod and battery

Battery and accessories

The mobile 50 Ah / 13.6 V LiPo battery in a Peli case powers the system where mains electricity is unavailable.

  • optional wide-FOV camera
  • spare cables and 12 V extension
  • transport case and maintenance set
Instrument footprint

Compact construction without a separate SpecBox

All measurement components fit inside one enclosure. The optical tube and connectors increase the overall installation envelope and must be considered when designing a mount or shelter.

300 × 200 × 132 mm instrument enclosure
123 mm long and 80 mm diameter optical tube
complete-system weight of approximately 7 kg
simple tripod, rail or mast mounting
AIRYX SkySpec Compact dimensions
Dimensioned views of the SkySpec Compact system; values are shown in millimetres.
Model applications

For short campaigns, plumes and mobile platforms

Compact is recommended primarily when a complete, portable system with automatic viewing-angle correction is required.

Point-source monitoringremote scanning of plumes from plants, power stations, stacks or volcanoes
Emission-rate retrievalcombining gas columns with wind data to calculate emission fluxes
Mobile measurementscars, ships and other moving platforms
Area budgetspollution inflow and outflow around cities and industrial areas
Short field campaignsmonitoring at remote locations with little infrastructure
Atmospheric researchMAX-DOAS sequences, gas columns and aerosol parameters

Ask about SkySpec Compact

We will select the RGD or HGD variant, target components, spectral range, software, power supply and field accessories.