AIRYX · SkySpec · UV/VIS spectrometer module

SpecBox Mini

A calibrated, characterised and actively temperature-stabilised grating spectrometer in a rugged enclosure. It is designed for integration with external optics, SkySpec telescopes and custom spectroscopic systems.

AIRYX SkySpec SpecBox Mini v.260SPECBOX MINI
<3 × 10⁻⁴noise at 1,000 scans
±0.03°Cspectrometer temperature stability
<10 selectronics start-up time
5 kgcomplete module weight
Design

Stable optics in a compact enclosure

SpecBox Mini combines a low-stray-light spectrometer, control electronics, temperature stabilisation and a fibre cross-section converter.

AIRYX SpecBox Mini design diagram

Prepared for precise DOAS analysis

The converter rearranges the circular fibre cross-section into a 7 × 100 µm slit geometry, improving throughput and spectrometer illumination uniformity.

factory wavelength calibration
offset and dark-current spectra
detector nonlinearity characterisation
CMOS or back-thinned CCD detector
external SMA fibre connection
Measured components

Trace gases and aerosol parameters

The final scope depends on detector, spectral configuration, external optics and analysis software.

NO₂nitrogen dioxide
O₃ozone
SO₂sulphur dioxide
HCHOformaldehyde
H₂Owater vapour
HONOnitrous acid
BrObromine monoxide
IOiodine monoxide
CHOCHOglyoxal
AODaerosol parameters
Technical data

SpecBox Mini v.260 specifications

Typical 300–460 nm configuration. Custom ranges, resolutions, detectors and fibres are available.

Spectral performance
TypeUV/VIS/IR grating spectrometer optimised for UV/VIS DOAS measurements
Typical range300–460 nm; custom configurations on request
Resolutionbelow 0.7 nm FWHM
Colour filterBG3
Noisebelow 3 × 10−4 at 103 scans, approximately 60 s integration
Spectral samplingmore than 6 points across the slit-function FWHM
Quantum efficiencyUV above 50%; VIS approximately 80%, detector dependent
F-numberf/4
Wavelength calibrationstable factory calibration; typical spectral shift below 0.01 nm
Characterisationdark current, offset, wavelength calibration and detector nonlinearity function
Fibre and detector
Internal fibrefused silica; circle-to-slit converter, 7 × 100 µm
External fibresingle 1 × 400 µm fibre with SMA connector; ordered separately
Detectorsregular-grade CMOS or UV-sensitive back-thinned CCD; other detectors on request
Stabilisation and operating conditions
Spectrometer temperature20°C, adjustable; stability better than ±0.03°C
Operating temperature−10 to +40°C
Sensorsmeasurement accuracy: temperature 1°C; pressure 0.5%; relative humidity ±3%
Condensation protectioninternal humidity monitoring and easily replaceable desiccant
Enclosurerugged IP64 polyethylene enclosure, except for the heat-exchanger fan
Control, power and mechanical data
Communicationone USB 2.0 port
SoftwareMS-DOAS for independent component control and measurement procedure creation
Start-up timebelow 10 s; full temperature stabilisation after approximately 10 min
Power supply12 V DC; typically below 25 W, 90 W maximum
Dimensions30 × 13.2 × 20 cm, width × depth × height; enclosure only
Weight5 kg
Results and equipment

From spectrum to DOAS fit

Spectral quality supports fit residuals of a few 10⁻⁴; longer exposures can further improve precision.

Example AIRYX SpecBox Mini DOAS fit

Example DOAS fit

The example with a four-minute exposure shows the spectral fit and residuals with a magnitude of a few 10⁻⁴.

  • low noise
  • resolution below 1 nm
  • precision can be improved by extending exposure time
AIRYX SpecBox Mini dimensions

Integration options

The module can operate with a SkySpec telescope or the customer’s external optics.

  • fibres and cables up to 20 m
  • 50 Ah, 13.6 V LiPo battery in a Peli case
  • DOAS analysis package, spare parts and online support
Applications

A module for custom systems and specialised research

SpecBox Mini does not include a telescope or external fibre. It is a configurable spectrometer module for integration.

Passive remote sensingDOAS measurements of trace gases and aerosols with suitable external optics
SkySpec systemsoperation with AIRYX telescopes or custom optical systems
Surface reflectancespectral reflection measurements of materials and surfaces
Material scienceprecise UV/VIS spectroscopy in laboratory and field work
Biomedical spectroscopyintegration with specialised optical paths
Stationary monitoringdesigned for long-term operation and permanent installations
Mobile applications12 V supply, battery option and rugged compact enclosure
Solution selection

Configure your SpecBox Mini

We will select the spectral range, detector, fibre, external optics, software and power supply for the planned system.

AIRYXMES-EKOinstrument and measurement-system selectionmeseko@meseko.pl