AIRYX · greenhouse-gas emissions from wastewater

WGA020

Wastewater Gas Analyser

A complete system for long-term monitoring of greenhouse-gas emissions from biological wastewater treatment. It measures N₂O, CH₄, CO₂ and O₂ in real time above aerated basins and automatically calculates net greenhouse-gas emission fluxes.

AIRYX WGA020 Wastewater Gas Analyser with analyser and sampling-line heating power unitWGA020
✓
0.5 ppmN₂O limit of detection (3σ)
✓
0.3 mg/m²/hN₂O emission-rate sensitivity
✓
6–12 monthstypical service; annual sensor calibration
✓
1 or 2 hoodsmonitoring at one or two measurement points
Why measure

Emission measurement, not just concentration

The WGA020 measures sample-gas concentrations, background concentrations and gas flow, allowing it to calculate net greenhouse-gas emission fluxes automatically, relative to the sampling area.

AIRYX floating sampling hood in an aerated wastewater-treatment basin

From the wastewater surface to a net-emission result

The system also records:

  • ambient temperature,
  • atmospheric pressure,
  • sample flow measured by an internal sensor,
  • system-health parameters.

Time resolution is 1 s internally, with results averaged to 60 s.

1
Floating sampling hoodCollects gases released from the aerated surface and follows changing water levels.
2
Total-flow measurementAn internal flow sensor supplies the quantity required to calculate emissions.
3
Drying and stabilisationSample-air drying protects sensors from high humidity, while temperature stabilisation limits the effects of ambient changes.
4
Four-gas analysisNDIR measures N₂O, CH₄ and CO₂, while an electrochemical sensor measures O₂.
5
Automatic net resultSample and background concentrations, flow and area are converted into an emission flux.
Analytical scope

Four gases, one process view

Simultaneous greenhouse-gas and oxygen measurements relate emissions to aeration conditions and biological changes in the basin.

N₂O0–1,000 ppmLOD 0.5 ppm; flux sensitivity 0.3 mg/m²/h.
CH₄0–1,000 ppmLOD 5 ppm; flux sensitivity 1.1 mg/m²/h.
CO₂0–10 vol.%LOD 40 ppm; flux sensitivity 24 mg/m²/h.
O₂0–100 vol.%LOD: 0.1 vol.% absolute or 0.5% of the reading.
Emission-rate sensitivity refers to a flow of 1 m³/h/m². Actual performance depends on flow, installation conditions and sampling configuration.
Data quality

Stable performance in wastewater conditions

The system addresses the principal challenges of humid-gas measurement above wastewater: condensation, drift, CO₂ interference and changing ambient temperature.

GasRangeLOD (3σ)Emission-rate sensitivity (σ)
N₂O0–1,000 ppm0.5 ppm0.3 mg/m²/h
CO₂0–10 vol.%40 ppm24 mg/m²/h
CH₄0–1,000 ppm5 ppm1.1 mg/m²/h
O₂0–100 vol.%0.1 vol.% absolute or 0.5% of readingn/a
Calibration and maintenance

The WGA020 is a low-maintenance system. Automatic internal corrections do not replace external sensor calibration, which is performed manually once a year using standard gases.

Typical servicing takes place every 6–12 months. Annual maintenance includes sensor calibration, scrubber refill or filter replacement, and sampling-hood cleaning.

Sampling

Floating sampling hood

The stainless-steel hood floats on the tank surface and follows dynamic water-level changes, providing stable gas collection above the biological basin. Its sampling area is approximately 0.3 m². Sampling-line length and construction are selected for the site.

Standard sampling line

Length 3–7 m, typically 5 m; 25 mm outer diameter.

Heated sampling line

75 mm outer diameter; self-regulated heating prevents condensation in cold climates.

Optional two-point monitoring

An optional configuration with two sampling hoods enables monitoring at two points. Measurements are quasi-simultaneous rather than fully simultaneous.

WGA020 floating hood with standard and heated sampling lines
Approximately 0.3 m² sampling area; optional dual-hood configuration with 2 × 0.3 m².
AIRYX WGA020 analyser rack, heating unit and sampling-hood dimensions
Mounting rack 1,540 × 500 × 340 mm; sampling hood 890 × 760 × 240 mm.
Site preparation

Installation requirements

The sampling-hood location and system arrangement are selected for the basin geometry, railing layout, sampling-line route and available utilities. Installation requires prior verification of site conditions.

Installation location

A safe, accessible location is required for the analyser and sampling hood. Selection requires photos of the basin, its edge and railings, the railing diameter and the basin dimensions.

Power supply

Installation requirements specify a 230 V AC supply through an IP67 CEE connection, preferably within 10 m. The power-supply configuration is confirmed during system selection.

Background air

A suitable pressurised-air supply is preferred; alternatively, a representative ambient-air sampling point can be used.

Communications

Ethernet and the web interface support local operation. Data can be stored on USB in UTF-8 text format. Remote LTE/VPN access is optional.

Connection details, sampling-line routing, background-air supply and communications are confirmed during system selection and in the site-specific quotation.
Technical data

Complete WGA020 specification

Parameters from the complete AIRYX WGA020 data sheet, issued 27 April 2026.

Measurement and calculations
Measured gasesN₂O, CO₂, CH₄, O₂
Measurement principleNDIR for N₂O, CO₂ and CH₄; electrochemical sensor for O₂; automatic cross-interference correction and sample-air drying
Measured parametersSample and background concentrations, total sample flow, greenhouse-gas net-emission fluxes, ambient temperature and pressure, and system-health parameters
Time resolution1 s internal; 60 s averages
Total sample flow0–20 slm
Analysed sample flowapproximately 0.2 slm
Background-air flow0.3–0.5 slm
Automatic internal correctionsInternal CO₂ cross-interference correction and automatic zero-drift correction
External calibrationAnnual manual sensor calibration using external standard gases
Operating conditions and construction
Operating temperature−10°C to +40°C
Ambient pressure800–1,150 hPa
EnclosureRugged, weatherproof design; IP64
Power supply220/230 V AC; the supply appropriate to the ordered configuration is confirmed in the quotation
Analyser power consumption80 W nominal; 120 W peak
Sampling-line heating0–156 W depending on temperature and length; typically 80 W at 10°C with a 5 m heated line
DimensionsMounting rack: 1,540 × 500 × 340 mm; sampling hood: 890 × 760 × 240 mm
WeightRack including instrument and heating unit: approximately 35 kg; sampling hood: approximately 8 kg; additional equipment: approximately 5 kg
Sampling
Sampling areaApproximately 0.3 m²; optional 2 × 0.3 m²
Sampling line3–7 m, typically 5 m; 25 mm outer diameter or 75 mm for the heated line
Background airPressurised air is preferred; alternatively, ambient air from a suitably selected reference location
Control and data
User interfaceTouch display; web interface from a connected PC or over the internet
Data interfacesEthernet; USB data export; optional remote VPN access through an internal LTE modem
Data formatUTF-8 text files
ServiceService typically every 6–12 months; annual sensor calibration, scrubber or filter servicing and sampling-hood cleaning
Data example

Dynamics that periodic sampling cannot reveal

High time resolution reveals daily peaks and short emission episodes. This is especially important for N₂O, whose climate impact is substantial even when concentration changes are small.

Example CO₂, N₂O and CH₄ emission fluxes measured by the AIRYX WGA020
Example field measurement of CO₂, N₂O and CH₄ fluxes expressed as CO₂ equivalents. Distinct daily peaks demonstrate process variability.
Applications

From emission inventories to process control

The WGA020 supplies data for climate reporting, aeration optimisation and research into biological process dynamics.

N₂O monitoringtracking emissions associated with nitrification and denitrification
Process optimisationrelating emissions to aeration, load and operating conditions
GHG reportinglong-term inventories of net emission fluxes
Compliance and reportingdata for monitoring requirements and wastewater-emission documentation
Two-point comparisonoptional dual hoods for spatial-variability assessment
Scientific researchconcentration and flux data for modelling basin processes
Long-term monitoringautonomous field operation with low service requirements
System selection

Configure the WGA020 for your treatment plant

We will define the mounting location, number of hoods, sampling line length and heating, background-air source, power, communications, commissioning and periodic service scope.

AIRYX MES-EKO measurement-instrument and system selection meseko@meseko.pl