Dual-beam NDIR
Dual-beam infrared absorption measurement for CH₄ and CO₂.
BioBasic Biomethane measures CH₄, CO₂, O₂ and H₂S in biomethane after upgrading and at grid injection. Continuous measurement at one-second intervals supports quality monitoring of the upgraded gas.
BIOBASIC BIOMETHANEBioBasic Biomethane monitors product-gas quality after biogas treatment, directly at the transfer point to the gas network. It combines CH₄, CO₂, O₂ and H₂S measurements to show the main biomethane components as well as oxygen and trace hydrogen sulphide.
Measurements every second quickly reveal quality fluctuations and support adjustments to the upgrading process. The analyser is a process-close complement to the gas chromatograph, whose readings are available every few minutes.
Dual-beam infrared absorption measurement for CH₄ and CO₂.
A galvanic oxygen sensor with temperature compensation from 5 to 45°C.
Optical H₂S measurement with a reference detector, covering 0–100 ppm.
Continuous readings help operators respond to changes in gas composition before transfer to the network.
CH₄ and CO₂ measurements show product-gas quality after the biogas upgrading stage.
O₂ and H₂S channels complement biomethane purity monitoring at the transfer point.
One-second readings show fluctuations between successive results from the process gas chromatograph.
Operation and data transmission are designed for continuous plant monitoring.
Gas-quality measurements in biomethane treatment and transfer systems.
| CH₄ / CO₂ | CH₄: 90–100 vol.%; CO₂: 0–10 vol.%; dual-beam NDIR |
|---|---|
| O₂ | 0–30 vol.%; electrochemical galvanic sensor |
| H₂S | 0–100 ppm; NDUV with reference detector |
| Resolution | CH₄: 0.1 vol.%; CO₂: 0.01 vol.%; O₂: 0.05 vol.%; H₂S: 0.1 ppm |
| Measurement | Every 1 s, continuous |
| Pressure compensation | 0.7–1.1 bar |
| Flow / pump | Internal pump: 4 l/min; optional pump-free version with supply pressure ≥60 mbar |
| Gas path / connections | PTFE; CK8/6 |
| Sample and calibration inputs | 1 sample input, 1 zero-gas input, 1 test-gas input |
| NDIR/NDUV cell temperature | Stabilised at 45°C |
| Ambient conditions | 5–45°C; approximately 90% relative humidity; a dehumidified indoor location is recommended |
| Housing | Wall-mounted, 400 × 600 × 220 mm; IP54; 18 kg |
| Display | Monochrome 128 × 64 pixel, 2.5-inch touch display |
| Standard interface | Isolated 4–20 mA outputs |
| Optional interfaces | Profibus DP, Profinet, Ethernet/IP, Modbus TCP |
| Power supply | 230 V AC / 50 Hz, 0.9 A, 207 VA; optional 115 V AC / 60 Hz or 230 V AC / 60 Hz |
The manufacturer specifies no additional conditioning for dry biomethane. Sample conditions must be considered when selecting the sampling point.
The internal pump provides 4 l/min. The optional pump-free version requires a sample supply pressure of ≥60 mbar.
Automatic zero/test-gas calibration takes approximately four minutes, during which no measurement results are provided. The calibration interval is configurable.
GA220 Biomethane is a simpler solution for continuous monitoring of the two main biomethane components.
See GA220 BiomethaneSend your measurement requirements to MES-EKO.