NDIR measurement
Methane and optional carbon-dioxide channels cover the main gas composition.
A plug-and-play online system for reliable biogas, digester-gas and landfill-gas quality measurements and process control.
BIOBASICA plug-and-play online system for reliable biogas, digester-gas and landfill-gas quality measurements and process control.
A plug-and-play stationary system for continuous biogas, digester-gas and landfill-gas quality measurement.
The measuring principle, gas path and sample conditioning are selected for the application.
Methane and optional carbon-dioxide channels cover the main gas composition.
H₂S, O₂ and optional H₂ channels supervise contaminants and process conditions.
The pump, condensate trap and monitored gas path form a compact measurement system.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Gas composition helps assess digester stability and conversion.
Combined channels support CHP, upgrading and landfill-gas operation.
Analogue and digital outputs provide values and status information.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Measured components | CH₄ and H₂S standard; optional CO₂, O₂ and H₂ |
|---|---|
| Ranges | CH₄ 0–100 vol.%; CO₂ 0–100 vol.%; O₂ 0–30 vol.% |
| H₂S / H₂ | H₂S 0–2,000 ppm standard or 0–10,000 ppm option; H₂ 0–10,000 ppm |
| Accuracy | ≤3% of measured value |
| Sample inputs | 1 standard; expandable to 5 |
| Pump | Approx. 4 l/min |
| Housing | 400 × 400 × 220 mm, IP54; approx. 18 kg |
| Outputs | RS232 and 4 × 4–20 mA; options available |
| Safety | ATEX-certified flame arresters; internal methane threshold monitoring |
| Long-term drift | Below 3% per year for IR channels |
| Measurement interval | 2 min standard; configurable from 2 to 999 min |
| Sample pressure | 0.7–1.1 bar, compensated |
| Gas path | PTFE |
| Power supply | 230 VAC / 50 Hz; optional 115 VAC / 60 Hz or 230 VAC / 60 Hz |
Key information for design, commissioning and subsequent analyser operation.
One input is standard and the system can be expanded for several plant locations.
The trap and level monitoring must be inspected as part of routine maintenance.
Flame arresters and internal methane supervision must remain operational.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.