Thermal-conductivity detector
The sensor compares heat transfer in the sample with the configured reference condition.
A multi-gas thermal-conductivity analyser for continuous process measurements where components have sufficiently different thermal conductivities.
FTC300A multi-gas thermal-conductivity analyser for continuous process measurements where components have sufficiently different thermal conductivities.
A multi-gas analyser based on thermal-conductivity detection for binary and application-specific gas mixtures.
The measuring principle, gas path and sample conditioning are selected for the application.
The sensor compares heat transfer in the sample with the configured reference condition.
Calibration accounts for the thermal conductivities of all relevant mixture components.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
The method supports gases that are difficult to measure optically.
Continuous data help supervise mixing, purity and conversion.
Digital interfaces support automation and data acquisition.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Principle | Thermal conductivity detection (TCD) |
|---|---|
| Measured components | H₂, CH₄, CO₂, O₂, NH₃, H₂S and further gases depending on matrix |
| Operating mode | Single- or multi-gas configuration |
| Interfaces | Modbus TCP, Ethernet and RS232 |
| Configuration | Ranges, accuracy and cross-sensitivity correction depend on the binary or multi-component gas matrix |
| Ranges | Up to 0–100%, depending on gas and component combination |
| Accuracy | Measuring-range start from approx. 0.5%, configuration-dependent |
| Operating temperature | −10 to +45 °C |
| Display | Touch display |
| Power supply | 24 VDC or 100–240 VAC |
| Readiness | Approx. 5 min; full specification after approx. 30 min |
Key information for design, commissioning and subsequent analyser operation.
All significant mixture components must be known during configuration.
Ranges and cross-sensitivity correction are established for the target composition.
Major matrix changes may require a revised calibration model.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.