Single-beam NDIR
Infrared absorption provides continuous and selective measurement of configured solvent vapours.
A continuous NDIR analyser for reliable supervision of harmful solvent concentrations in environmentally and safety-critical processes.
GA 150A continuous NDIR analyser for reliable supervision of harmful solvent concentrations in environmentally and safety-critical processes.
A continuous NDIR analyser for rapid and reliable supervision of harmful solvents in environmentally and safety-critical processes.
The measuring principle, gas path and sample conditioning are selected for the application.
Infrared absorption provides continuous and selective measurement of configured solvent vapours.
The 60 °C measuring cell helps retain less volatile compounds in the gas phase.
Sample-pressure effects are compensated within the permitted inlet range.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Continuous measurement detects process leakage and concentration changes.
Trend and alarm signals can support protection against hazardous solvent exposure.
Stable data help control cleaning, surface-treatment and solvent-based processes.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Measured components | PCE, TCE, DCE and other selected IR-active solvents |
|---|---|
| Principle | Single-beam NDIR; cell heated to 60 °C |
| Type | Continuous online measurement |
| Range | Approx. 2 mg/m³ to 40 g/m³, component-dependent |
| Accuracy | Below 3% of measured value |
| Sample inputs | 1 standard; expandable to 8 |
| Flow / pressure | Approx. 4 l/min; 0.7–1.1 bar |
| Housing | 19-inch rack |
| Interfaces | RS232; optional industrial Ethernet/fieldbus interfaces |
| Ambient conditions | 5–45 °C |
| Display | 2.5-inch resistive touch display |
| Power supply | 230 VAC / 50 Hz, 0.9 A, 207 VA |
| Readiness | Initial results after approx. 60 s; full specification after 30 min |
| Gas-path option | PTFE standard; stainless steel optional |
| Sample-point capacity | One standard; expandable to eight |
Key information for design, commissioning and subsequent analyser operation.
Range and optical configuration are selected for the target solvent and matrix.
PTFE or stainless steel is selected according to chemical compatibility.
Initial values are available quickly; full specification applies after approximately 30 minutes.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.