Paramagnetic measurement
Oxygen is attracted by a magnetic field, enabling selective concentration measurement without chemical consumption of the cell.
A continuous oxygen analyser with a paramagnetic measuring cell, short response time and high long-term stability.
GA 210 O₂A continuous oxygen analyser with a paramagnetic measuring cell, short response time and high long-term stability.
A continuous oxygen analyser using the paramagnetic properties of oxygen. The measuring cell provides stable O₂ supervision without a chemically consumed electrolyte.
The measuring principle, gas path and sample conditioning are selected for the application.
Oxygen is attracted by a magnetic field, enabling selective concentration measurement without chemical consumption of the cell.
Controlled flow and effective moisture removal protect measurement stability.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
The absence of a consumed electrolyte limits changes in sensor characteristics over time.
Configurable sample inputs and interfaces support analyser-rack and automation integration.
Continuous O₂ data support supervision of process atmospheres, combustion and gas purity.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Measured component | O₂ |
|---|---|
| Principle | Paramagnetic measuring cell |
| Ranges | 0–25, 0–50 or 0–100 vol.% O₂ |
| Resolution | 0.1 vol.% |
| Measurement interval / response | 1 s / below 10 s |
| Sample inputs | 1 standard; expandable to 6; two calibration-gas inputs |
| Flow | Approx. 4 l/min inlet; approx. 100 ml/min through cell |
| Pressure | 0.7–1.1 bar |
| Housing / weight | 19-inch, IP20; 13 kg |
| Warm-up | First values after approx. 5 min; full specification after 30 min |
| Accuracy / stability | ±0.1 vol.% O₂ per week; zero deviation below ±0.1 vol.% O₂ per week |
| Sample conditioning | Additional gas drying required |
| Ambient conditions | 5–45 °C; up to approx. 90% relative humidity, non-condensing |
| Display / power | 2.5-inch 128 × 64 touch display; 230 VAC / 50 Hz, 207 VA |
Key information for design, commissioning and subsequent analyser operation.
Initial measurement is available after approximately five minutes; full specification applies after approximately 30 minutes.
The inlet flow is approximately 4 l/min, with about 100 ml/min through the cell.
The sample must be dried and supplied within the conditions specified for the configuration.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.