Single-beam NDIR
Measurement of infrared-active gases. Automatic ZeroLock zero-point correction and cross-sensitivity compensation reduce drift and calibration effort.
A flexible continuous analyser platform using NDIR, electrochemical or zirconium-oxide measurement principles depending on the target component.
GA 210Gas Analyzer 210 is a flexible platform for continuous gas measurement in industrial installations. Depending on the measuring task, it can use single-beam NDIR, an electrochemical sensor or a zirconium-oxide measuring cell. The analyser can therefore be configured for solvent monitoring as well as oxygen, hydrogen or hydrogen-sulphide measurement.
The platform is designed for process control, environmental protection and plant-safety applications. Typical fields include the chemical and pharmaceutical industries, recycling, waste treatment, energy generation and exhaust-gas analysis.
In the NDIR version, the analyser measures infrared-active gases including perchloroethylene (PCE), trichloroethylene (TCE), dichloromethane, R11, R12, cyclopentane and other hydrocarbons. Oxygen can be measured with an electrochemical or ZrO₂ cell, while electrochemical sensors are also available for components such as H₂S and H₂.
The rear-panel view helps assess sample connections, communication interfaces and system integration.


The measuring principle, gas path and sample conditioning are selected for the application.
Measurement of infrared-active gases. Automatic ZeroLock zero-point correction and cross-sensitivity compensation reduce drift and calibration effort.
A cost-effective measuring principle for components including O₂, H₂S and H₂, particularly at lower concentrations and under controlled sample conditions.
An oxygen-measurement principle suitable for elevated temperature and humidity, providing fast response and a long sensor service life.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Continuous and stable results help identify deviations earlier and allow process parameters to be corrected more quickly, reducing rejected batches, raw-material consumption and energy losses.
Open communication interfaces and configurable 4–20 mA outputs simplify connection to the existing control system, data acquisition and alarm logic.
Measurement of critical components supports limit-value supervision, emissions documentation and the plant response to hazardous concentrations. Alarm functions are configured for the individual project.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Sample inputs | 1 as standard; optionally expandable to 8 |
|---|---|
| Measurement interval | 1 s in continuous mode or 20 s in quasi-continuous mode |
| Measuring principles | Single-beam NDIR; electrochemical sensor; zirconium oxide ZrO₂ |
| Example NDIR components | PCE, R11, R12, cyclopentane and many other infrared-active gases |
| Example EC components | O₂, H₂S and H₂ |
| ZrO₂ component | O₂ |
| Measuring ranges | Up to 100 vol.% or 0–10,000 ppm, depending on the component and configuration |
| Accuracy | Dependent on the measured component and selected range |
| Standard sample flow | 4 l/min |
| Permissible sample pressure | 0.7–1.1 bar |
| Gas path | PTFE or stainless steel |
| Display | 128 × 64 mm colour touch display |
| Digital interfaces | RS232; optional Profibus DP, Ethernet/IP, Modbus TCP and Profinet |
| Analogue outputs | Optionally up to 8 × 4–20 mA |
| Warm-up time | Full compliance with the specification after 30 minutes |
| Power supply | 230 VAC / 50 Hz, 0.9 A, 207 VA; 115 VAC / 60 Hz and 230 VAC / 60 Hz versions |
| Housing and protection rating | 19-inch rack, 400 mm depth; IP20 |
| Weight | 13 kg |
Key information for design, commissioning and subsequent analyser operation.
The analyser is a single-component instrument. It has one sample input as standard and can be expanded to eight inputs to measure the same component sequentially at different points in the installation.
Initial measured values may be available earlier, while full compliance with the declared performance is reached after approximately 30 minutes of warm-up.
Long-term stable sensors and automatic ZeroLock correction in the NDIR version reduce calibration effort. The maintenance schedule is selected for the measured gas, sample conditions and the plant's quality requirements.
RS232 is standard. Options include Profibus DP, Ethernet Modbus TCP, Ethernet/IP and Profinet, as well as up to eight configurable and galvanically isolated 4–20 mA outputs.
Configurations for gases other than those listed are possible. Each application requires verification of the sample properties, expected range and potential cross-sensitivities.
Stable measurement helps detect deviations early and reduce rejected production, downtime and unnecessary energy or raw-material consumption. The data can also support emissions supervision and designed alarm functions.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.