Dual-beam NDIR
Dual-beam CH₄ and CO₂ measurement with the measuring cell heated to 50°C.
GA220 Biomethane is a CH₄ and CO₂ analyser for continuous biomethane monitoring after upgrading and at grid injection. Dual-beam NDIR measures every second across CH₄ 90–100 vol.% and CO₂ 0–10 vol.%.
GA 220 BIOMETHANEGA220 Biomethane is a dedicated variant for CH₄ and CO₂ measurement in upgraded biomethane. CH₄ 90–100 vol.% and CO₂ 0–10 vol.% ranges support continuous composition monitoring after upgrading, with a reading every second.
Dual-beam NDIR, pressure compensation and high long-term stability reduce calibration effort. The open interface architecture supports transmission of measured values to plant automation systems.
Dual-beam CH₄ and CO₂ measurement with the measuring cell heated to 50°C.
Measurement accounts for process-gas pressure in the 0.7–1.1 bar range.
Two measurement channels provide current data on the main components of upgraded biomethane.
The high-CH₄ range and residual CO₂ measurement are designed for biomethane quality monitoring.
Continuous readings track concentration changes without waiting for a longer measurement cycle.
RS232 and optional industrial protocols transmit measured values to the control system.
Measurement and operation features reduce the instrument’s operating requirements.
CH₄ and CO₂ measurement in biomethane plants and other applications listed for this variant.
| CH₄ | 90–100 vol.% |
|---|---|
| CO₂ | 0–10 vol.% |
| Method | Dual-beam NDIR; measuring cell heated to 50°C |
| Measurement interval | 1 s |
| CH₄ resolution / detection limit | 0.1% of full scale (FS) |
| CO₂ resolution / detection limit | 0.05% of full scale (FS) |
| Expanded uncertainty | CH₄: 0.2% FS; CO₂: 0.5% FS |
| Pressure compensation | 0.7–1.1 bar |
| Flow / pump | 4 l/min; optional pump |
| Gas path | PTFE as standard; stainless steel optional |
| Filter | Sintered stainless-steel filter |
| Connections | CK6/4; alternatively 6/4 cutting-ring fittings |
| Housing | 19-inch rack, 400 mm depth; IP20; 12 kg |
| Standard interface | RS232 |
| Optional interfaces | Profibus DP, Modbus TCP, Ethernet/IP, Profinet; up to 8 isolated 4–20 mA outputs |
| Measurement readiness | Approx. 5 min; full specification after approx. 30 min |
| Ambient conditions | 5–45°C; approximately 90% relative humidity, non-condensing |
| Power supply | 230 V AC / 50 Hz, 0.9 A, 207 VA; optional 115 V AC / 60 Hz or 230 V AC / 60 Hz |
Measurement readiness is reached after approximately 5 min; full specification is achieved after approximately 30 min.
Standard flow is 4 l/min. A pump is optional; the standard PTFE gas path can optionally be made from stainless steel.
RS232 is standard. Industrial protocols and up to eight isolated 4–20 mA outputs are optional.
BioBasic Biomethane extends continuous CH₄ and CO₂ measurement with O₂ and H₂S monitoring.
See BioBasic BiomethaneSend your measurement requirements to MES-EKO.