Multi-channel NDIR
Infrared absorption channels measure selected IR-active gases continuously.
Developed for quick and easy online measurements of up to six process gases where mutual interferences must be compensated and low detection limits are required.
GA 220Developed for quick and easy online measurements of up to six process gases where mutual interferences must be compensated and low detection limits are required.
A stationary multi-channel platform for continuous process and emissions gas monitoring. NDIR channels can be combined with additional sensors and application-specific sample handling.
The rear-panel view helps assess sample connections, communication interfaces and system integration.


The measuring principle, gas path and sample conditioning are selected for the application.
Infrared absorption channels measure selected IR-active gases continuously.
Electrochemical or zirconium-oxide channels can extend the analyser to non-IR components.
Sample-pressure effects are corrected to improve result stability.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Continuous measurement supports limit supervision and plant documentation.
Early detection of deviations enables faster process correction.
Several channels provide a common view of the gas composition.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Measurement channels | 0–3 NDIR channels plus additional sensor channels depending on configuration |
|---|---|
| NDIR cell | Single-beam, heated to 50 °C |
| Example components | PCE, R11, R12, cyclopentane and other IR-active gases |
| Range / accuracy | Gas- and configuration-dependent |
| Sample flow / pressure | Approx. 4 l/min; 0.7–1.1 bar with compensation |
| Gas path | PTFE or stainless steel depending on medium |
| Power | 230 VAC / 50 Hz, 207 VA |
| Housing / weight | 19-inch, 400 mm depth, IP20; approx. 13 kg |
| Interfaces | RS232 standard; industrial interfaces and analogue outputs optional |
| Sample outlet | 1 |
| Measured gases | 1–3 NDIR plus 1–3 EC components; up to 6 gases depending on configuration |
| Measurement interval | Dependent on component and configuration |
| Measuring technologies | Single-beam NDIR, electrochemical sensors and optional ZrO₂ for O₂ |
| Measuring-cell temperature | 50 °C |
| Measuring range | Selected individually for the component and application |
| Accuracy / detection limit | Dependent on component, range and optical configuration |
| Compensation | Pressure and cross-sensitivity compensation; up to three IR interferents |
| Sample conditioning | Optional sample filter and pressure reducer for inlet pressure up to 100 bar |
| Internal filter | Sintered stainless-steel filter |
| Ambient conditions | 5–45 °C; up to approx. 90% RH, non-condensing |
| Display / operation | 128 × 64 monochrome display; 2.5-inch resistive touch panel |
| Digital inputs | 4 × 24 V AC/DC |
| Digital outputs | Ready and fault relays plus four configurable outputs; max. 48 V / 500 mA |
| Analogue outputs | Optionally up to eight galvanically isolated 4–20 mA outputs |
| Power supply | 230 VAC / 50 Hz, 0.9 A, 207 VA |
| Readiness | Component-dependent; full specification after 30 min |
Key information for design, commissioning and subsequent analyser operation.
Optical path length and range are selected for each component and expected concentration.
Temperature, humidity, pressure and potential interferences must be defined.
Results can be transferred to a higher-level system through industrial interfaces.