NDIR for formaldehyde
The optical system is optimised for the characteristic absorption of HCHO.
An online NDIR analyser for formaldehyde and other IR-active components in hot, humid gas streams where conventional gas cooling may cause losses.
GA 315 FORMALDEHYDEAn online NDIR analyser for formaldehyde and other IR-active components in hot, humid gas streams where conventional gas cooling may cause losses.
A specialised heated NDIR analyser for formaldehyde measurement directly in hot, humid exhaust gas, including gas-engine applications. The heated gas path limits HCHO losses caused by cooling and condensation.
The measuring principle, gas path and sample conditioning are selected for the application.
The optical system is optimised for the characteristic absorption of HCHO.
The heated cell and sample path keep formaldehyde in the gas phase.
The configured measurement compensates relevant effects from methane, humidity and sample pressure.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Measurement supports formaldehyde-emissions assessment from combined heat and power units.
Avoiding sample cooling reduces HCHO losses to condensate.
Emission trends help assess engine settings and fuel-gas quality.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Measured component | Formaldehyde (HCHO); configuration-specific additional IR-active components |
|---|---|
| Principle | Heated NDIR measurement for hot, humid sample gas |
| Formaldehyde range | 0–160 ppm in the supplied brochure configuration |
| Detection limit | 3 ppm formaldehyde |
| Compensation | Methane, humidity and sample-pressure compensation |
| Pressure | 0.7–1.1 bar |
| Power | 230 V / 50 Hz; 1300 VA |
| Interface | RS232 standard; project-specific outputs/interfaces |
| Accuracy | Below 5% of measured value |
| Sample input | One heated gas input on the front panel |
| Sample temperature | Components in gas streams up to 300 °C with suitable heated sampling |
| Sample flow | Approx. 4 l/min |
| Gas path | Stainless steel with compression fittings |
| Internal filter | 90 µm sintered stainless-steel filter |
| Sample connection | Quick coupling for heated analytical line |
| Ambient conditions | 5–45 °C; up to approx. 90% RH, non-condensing |
| Display | 840 × 640 pixel touch panel |
| Digital inputs | 6 × 24 V AC/DC |
| Digital outputs | One ready relay and four relay outputs; max. 48 V / 500 mA |
| Analogue outputs | Four galvanically isolated 4–20 mA outputs |
| Housing | 19-inch rack, 400 mm depth, IP20 |
| Warm-up time | Approx. 30 min at 20 °C ambient |
Key information for design, commissioning and subsequent analyser operation.
The complete path to the cell must remain above the condensation point.
Humidity and potential interfering components must be defined before selection.
Heated lines and filters require periodic inspection for contamination and free flow.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.