Surface-gas collection
The hood isolates a defined liquid area and directs released gas to the sample outlet.
A floating hood for gas collection over a defined 1.00 m² wastewater-basin area. Gas Dome 300 is offered exclusively with Fresenius software — as part of the complete Off-Gas System or for technically agreed integration into an existing analysis system.
GAS DOME 300The complete Fresenius Off-Gas System is available in Smart and Expert versions. It combines Gas Dome 300, gas routing and conditioning, flow measurement, analyser and outdoor cabinet. Integral Fresenius software calculates N₂O, CH₄ and CO₂ emission rates.
View the complete Off-Gas System →
Fresenius software is an integral part of every Gas Dome 300 solution. The hood is not offered as a stand-alone product without this software.
Gas Dome 300, gas routing and conditioning, flow measurement, analyser, outdoor cabinet and emission-rate calculation form one coordinated solution.
Gas Dome 300 together with Fresenius software can be integrated with a suitable existing or third-party customer analyser. Equipment compatibility, interfaces, signals, measuring ranges and operating conditions must be agreed with Fresenius before integration.
The measuring principle, gas path and sample conditioning are selected for the application.
The hood isolates a defined liquid area and directs released gas to the sample outlet.
The PVC construction follows changes in basin water level.
Slight overpressure under the hood transports gas to the agreed analysis system without an additional blower. Measurement performance depends on the connected analyser and gas path.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Fresenius software combines concentration and flow measurements with the hood area to determine the emission rate.
Repeatable hood placement improves comparison between basin locations.
Integration with a suitable Fresenius or third-party analyser requires Fresenius software and prior technical coordination.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Function | Passive gas collection over a defined liquid surface area |
|---|---|
| Construction | Black pyramidal PVC hood |
| Gases | N₂O, CH₄, CO₂ — measured by the connected analyser |
| Collection area | Exactly 1.00 m² |
| Skirt | Immersion depth 20–30 cm |
| Gas hose | Internal diameter ID 40 mm; standard length 25 m per measuring point; laid with a gradient towards the hood |
| Gas transport | Slight overpressure under the hood; no additional blower |
| Position keeping | Guy ropes led to the basin edge |
| Gas path of the complete Off-Gas System | Condensate trap with level monitor, sample gas filter and flow measurement in a DN50 metering tube — a separate gas path for each point |
| Software | Fresenius software is integral and required in every solution; the hood is not supplied as a stand-alone product without software |
| Compatibility | Complete Fresenius system or suitable customer analyser with Fresenius software; interfaces, ranges and operating conditions agreed before integration |
| Measurement performance | Accuracy, range and detection limit are determined by the connected analyser, not by the hood |
| Emission-rate calculation | E = c × Q / A; c — mass concentration, Q — gas flow at standard reference conditions, A = 1.00 m²; result in mg/(m²·h) |
| Minimum flow for a valid emission result | 1 m³/h; below this level, the result must not be interpreted as a valid emission value |
| Two points in the complete Off-Gas System | One cabinet and one analyser serve up to two hoods; the second point requires a hood, hose and its own gas path |
The system determines the emission rate for a defined measurement area. It does not automatically replace an official total-emissions inventory or a CEMS.
Key information for design, commissioning and subsequent analyser operation.
Select a representative surface area and document the active collection area.
Ropes should stabilise the hood without preventing changes in liquid level.
Lay the hose with a gradient towards the hood; gas-tight connections, moisture and condensate must be considered when designing the gas path.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.