FRESENIUS · COMPLETE OFF-GAS SYSTEM

Measurement of N₂O, CH₄ and CO₂ emissions from wastewater treatment plants

Fresenius Off-Gas System combines Gas Dome 300, the gas path, an outdoor cabinet, an analyser and emission-rate calculation. It measures gases where they are released from an aeration tank and converts concentration and flow into an emission rate per unit area.

Fresenius Off-Gas System with Gas Dome 300 on an aeration tankOFF-GAS SYSTEM
N₂O / CH₄ / CO₂three measured components
1.00 m²Gas Dome 300 collection area
mg/(m²·h)automatic emission-rate calculation
no cloud or subscriptionyour data stays at the plant

How the Off-Gas System works

1

Gas Dome 300

The floating hood captures gas from 1.00 m² of aeration-tank surface.

2

Gas transport

Pressure under the hood transports the gas through a hose to the outdoor cabinet beside the tank. No additional blower is required.

3

Gas analysis

The analyser measures N₂O, CH₄ and CO₂ concentrations. Smart uses Smart Control 300, while Expert uses GA 320.

4

Emission-rate calculation

The system automatically calculates the emission rate in mg/(m²·h) and can transfer the data to the plant control system.

mg/(m²·h)

Concentration and emission rate

Concentration describes how much of a component is present in the gas. The emission rate combines concentration with measured flow and the known hood area of 1.00 m². The analyser firmware performs this calculation without additional external software.

Common to Smart and Expert

Gas Dome 300

Both versions share Gas Dome 300, gas routing, flow measurement, emission-rate calculation and the outdoor cabinet. The main difference lies in the analyser and measuring performance: Smart uses Smart Control 300 (SC300), while Expert uses GA 320.

View Gas Dome 300 →
Gas Dome 300 – Off-Gas System collection hood
Smart Control 300Smart Control 300

Off-Gas System Smart

Operational monitoring, routine emission assessment and continuous monitoring where concentrations in the two-digit ppm range are sufficient.

View Smart Control 300 →
GA 320GA 320

Off-Gas System Expert

Low emission levels, measurement campaigns and more demanding applications, including plants with optimised nitrogen removal.

View Gas Analyzer 320 →

Smart vs Expert – technical data

ParameterSmartExpert
AnalyserSmart Control 300 (SC300)GA 320
N₂O LoD10 ppmV0.5 ppmV
CH₄ LoD100 ppmV2 ppmV
CO₂ LoD0.04 vol.%1 ppmV
N₂O resolution1 ppmV0.1 ppmV
CH₄ resolution1 ppmV0.1 ppmV
CO₂ resolution0.01 vol.%0.1 ppmV
Expanded concentration measurement uncertainty≤10% of reading up to 30 ppmV N₂O or 300 ppmV CH₄; above these levels ≤5% of reading≤2% of reading per component
Reading correctionno humidity compensationhumidity and cross-sensitivity compensation
N₂O range0–2,000 ppmV0–2,000 ppmV
CH₄ range0–1,000 ppmV0–10,000 ppmV
CO₂ range0–5 vol.%0–10,000 ppmV
Optional O₂ measurementEC sensor; 0–30 vol.% (option)EC sensor; 0–30 vol.% (option)
Measuring interval21 s per inlet20 s per inlet
Analyser gas inlets2up to 6; 2 used in the system
Operation2.5″ display7″ touch display
Outdoor cabinet (H × W × D)1200 × 600 × 400 mm1200 × 600 × 600 mm
Power supply230 V / 50 Hz230 V / 50 Hz
Internal cabinet temperature during operation5–45°C5–45°C
Cabinet installationOutside any hazardous areaOutside any hazardous area
Weight ready for operation55–70 kg75–90 kg
Emission-rate uncertainty at 20 / 10 / 5 m³/h flow±11.1% / ±11.9% / ±14.2%±5.2% / ±6.8% / ±10.3%

Concentration measurement uncertainty and the complete system’s emission-rate uncertainty are different quantities. The ≤2% figure must not be interpreted as the uncertainty of the complete emission measurement.

Data from the Fresenius Off-Gas System Smart brochure dated 07.10.2026, the Expert brochure dated 06.10.2026 and the Smart/Expert comparison dated 06.10.2026.

Gas Dome 300 and the shared gas path

Gas Dome 300 areaExactly 1.00 m²
Hood skirtImmersion depth 20–30 cm
Hose per measuring pointID 40 mm; standard length 25 m, laid with a gradient towards the hood
Flow from the hoodSlight overpressure; no additional blower
Gas path per pointCondensate trap with level monitor, sample gas filter and thermal flow meter in a DN50 metering tube, range 0–21 m³/h
Emission-rate calculationE = c × Q / A; c — mass concentration, Q — gas flow at standard reference conditions, A = 1.00 m²; result in mg/(m²·h). The N₂O result can also be expressed as N₂O-N.
Minimum flow criterion1 m³/h; below this level the reading is flagged as invalid and must not be interpreted as a valid emission value
Measuring pointsCabinet and analyser prepared for two points; each hood has its own hose and gas path

Shared system construction

Both versions share a gas path with condensate trap, filter and DN 50 thermal flow measurement, and a heated, ventilated IP55 / IK10 outdoor cabinet beside the tank. The system supports one or two measuring points.

No zero or span gas

Both analysers use self-referencing NDIR with a dual-pressure reference and automatic zero-point correction at every measurement. During normal operation, they require no routine zero or span gas and no routine zero/span calibration.

Scope of supply and site requirements

The system is supplied assembled, wired and factory-tested, ready for connection to power and the control system.

Customer-side services include:

  • foundation / installation location,
  • electrical supply up to the cabinet,
  • connection to the control system beyond the terminal box,
  • required civil works at the basin.

Extension to a second measuring point

One cabinet and one analyser serve up to two hoods. The second point requires an additional Gas Dome 300, gas hose and its own gas path — a second complete system is not required.

The cabinet and analyser are prepared for two points. Each point has its own condensate trap, sample gas filter and DN50 flow measurement.

Integration with the wastewater plant control system

Smart / SC300: RS-232, analogue and relay outputs of the GA platform; optional Profibus DP, Modbus TCP, EtherNet/IP and Profinet.

Expert / GA320: RS-232, up to 8 analogue outputs 4–20 mA and relay outputs; optional Profibus DP, Modbus TCP, EtherNet/IP and Profinet.

The analyser calculates the emission rate internally. No cloud or subscription is required.

Measuring actual emissions at the treatment plant

According to Fresenius, the revised EU Urban Wastewater Treatment Directive increases the importance of accounting for CH₄ and N₂O emissions from wastewater treatment plants. The system enables direct measurement where emissions occur and can support operational monitoring, process assessment and documentation of actual emission levels.

The system determines the emission rate for a defined measurement area. It does not automatically replace an official total-emissions inventory or a CEMS.

Select Smart or Expert for your treatment plant

Provide the basin type, expected concentrations, number of points and measurement purpose: process monitoring, assessing changes or a measurement campaign.

Ask about configuration