Controlled gas cooling
The adjustable cooling block lowers sample temperature and condenses excess moisture.
Deliberately lowers the dew point of humid gases to protect the measurement system from condensation and stabilise the outlet dew point.
RCI 100Deliberately lowers the dew point of humid gases to protect the measurement system from condensation and stabilise the outlet dew point.
A compact sample-gas cooler for stable cooling and dehumidification of one gas stream before analysis.
The measuring principle, gas path and sample conditioning are selected for the application.
The adjustable cooling block lowers sample temperature and condenses excess moisture.
Separated condensate is removed from the gas path before the sample reaches the analyser.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
Controlled cooling reduces humidity-related measurement variation.
Moisture removal protects cells, filters and downstream tubing.
The housing can be incorporated into a complete conditioning system.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Function | Cooling and dehumidification of one sample-gas stream |
|---|---|
| Cooling block | Adjustable 2–20 °C |
| Maximum flow | 260 l/h |
| Cooling capacity | Up to 300 kJ/h |
| Housing | 225 × 280 × 305 mm, IP21 |
| Weight | 13.2 kg |
| Power | 230 VAC, approx. 150 VA |
| Interface | RS232 |
| Cooling technology | Compressor cooling |
| Cooling-block temperature | 5 °C standard; adjustable from 2 to 20 °C |
| Permissible sample pressure | 0.7–1.1 bar |
| Ambient temperature | 10–45 °C |
| Cooling path | Duran-glass cooling flask and condensate pump |
| Display / operation | 128 × 64 monochrome display; 2.5-inch resistive touch panel |
| Digital I/O | 4 × 24 V AC/DC inputs; ready, fault and four configurable relay outputs |
| Sample connections | CK 6/4 screw fittings; 6/4 compression fittings optional |
| Power / readiness | 230 VAC / 50 Hz, 150 VA; approx. 15 min readiness at 20 °C |
| Condensate removal | Integrated condensate pump |
Key information for design, commissioning and subsequent analyser operation.
Gas temperature, flow and moisture load must remain within the cooling capacity.
Drainage and pump operation require regular inspection.
Allow the cooling block to stabilise before starting measurement.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.