Two independent gas paths
Separate exchangers cool two sample streams without mixing them.
A 19-inch cooler for simultaneous conditioning of two sample-gas streams with stable and adjustable block temperature.
RCI 700A 19-inch cooler for simultaneous conditioning of two sample-gas streams with stable and adjustable block temperature.
A two-channel sample-gas cooler for independent conditioning of two gas streams in stationary analyser systems.
The measuring principle, gas path and sample conditioning are selected for the application.
Separate exchangers cool two sample streams without mixing them.
The cooling block provides a stable outlet condition for downstream analysers.
Measurement results can support process optimisation, reporting, quality control and designed alarm functions.
One cooler conditions two independent measurement streams.
Stable cooling protects analysers against moisture carry-over.
The 19-inch format and communication options support analyser cabinets.
The final configuration depends on gases, ranges, sample conditions and system integration.
The range and final configuration are confirmed for the medium and process conditions.
| Function | Cooling of two independent sample-gas streams |
|---|---|
| Gas paths | 2 inputs / 2 outputs |
| Maximum flow | 260 l/h per input |
| Cooling capacity | Up to 300 kJ/h |
| Housing | 19-inch slide-in, 484 × 405 × 350 mm, IP21 |
| Weight | 15.5 kg |
| Power | 230 VAC, approx. 150 VA |
| Interface | RS232 |
| Readiness | Approx. 15 min |
| Cooling technology | Compressor cooling |
| Cooling-block temperature | 5 °C standard; adjustable from 2 to 20 °C |
| Permissible sample pressure | 0.7–1.1 bar |
| Ambient temperature | 10–45 °C |
| Cooling path | Duran-glass cooling flasks and condensate pump |
| Display / operation | 128 × 64 monochrome display; 2.5-inch resistive touch panel |
| Digital I/O | 4 × 24 V AC/DC inputs; ready, fault and four configurable relay outputs |
| Sample connections | CK 6/4 screw fittings; 6/4 compression fittings optional |
| Condensate removal | Integrated condensate pump for both cooling paths |
Key information for design, commissioning and subsequent analyser operation.
Each inlet must remain within the stated maximum flow.
Both drainage circuits require inspection and preventive maintenance.
Approximately 15 minutes should be allowed for thermal stabilisation.
Provide the key measurement information. MES-EKO will check the instrument configuration and required sampling components.