GASERA · hydrogen fluoride

GASERA ONE HF

A highly selective hydrogen fluoride analyser for continuous monitoring of trace HF concentrations in clean rooms, ambient air and industrial installations.

GASERAONE HF
GASERA ONE HF — product view 2ONE HF
HF
Sub-ppb measurement capabilityFast response to concentration spikes
<1
No consumables or wet chemistryLow sample volume
10
Fence-line and ambient HFdo 10 ppm
24
Aluminium smelters and refineriespraca autonomiczna 24/7
0
Petrochemical industryOccupational safety
12
High-temperature processesHF
Why does HF require special attention?

Toxic, corrosive and exceptionally difficult to sample

Hydrogen fluoride forms highly corrosive hydrofluoric acid on contact with moisture. It is also a sticky molecule, so wetted materials and a low cell volume are crucial.

4 Aluminium smelters and refineries. Final performance depends on the selected gases, sample matrix and operating conditions.
ppb Petrochemical industry. Final performance depends on the selected gases, sample matrix and operating conditions.
HF + H₂O Occupational safety. Final performance depends on the selected gases, sample matrix and operating conditions.
„sticky” High-temperature processes. Final performance depends on the selected gases, sample matrix and operating conditions.
Key benefits

A solution designed specifically for HF

A small gold-coated measurement cell, carefully selected materials and a laser tuned to an HF absorption line address the challenges typical of this reactive gas.

01

Sub-ppb LoD

The standard configuration achieves a detection limit below 0.5 ppb; a higher-sensitivity variant can reach approximately 0.2 ppb.

02

High selectivity

The tunable diode laser operates on an isolated hydrogen fluoride absorption line.

03

Low-volume cell

A short path and low volume reduce HF adsorption and improve system response.

04

Stable operation

Stabilised cell temperature and pressure reduce the influence of changing ambient conditions.

05

No consumables

The optical measurement requires no reagents, cartridges or wet chemistry.

06

Field operation

The 19-inch housing can be used on a bench, in a rack or in a portable transport enclosure.

Example applications

From clean rooms to heavy industry

ONE HF is suited to applications where rapid detection of low hydrogen fluoride levels protects people, processes and product quality.

🧬

Microelectronics

Monitoring clean rooms and microfabrication processes in which HF is a key reagent.

🏭

Aluminium industry

Monitoring HF emissions from high-temperature processes and aluminium smelting installations.

🛢️

Petrochemicals

Detection of process leaks and exposure monitoring around process installations.

🧪

Fluorochemical production

Emission monitoring during the production of fluorides, fluorocarbons, polymers and pharmaceuticals.

🌾

Phosphate fertilisers

Measurement of HF emissions and monitoring around fertiliser plants.

♻️

Hazardous waste

HF measurement in waste streams and around waste-treatment facilities.

🛡️

Occupational health and safety

Early detection of releases, leaks and hazardous workplace concentrations.

🌫️

Ambient air

Long-term monitoring of low HF levels around industrial plants.

Measurement technology

CEPAS + tunable diode laser

The measurement uses a strong, isolated HF absorption line in the near infrared and a patented optical MEMS cantilever sensor.

1
Detection limit<0.5 ppb at 60 s
2
Range0–10 ppm
3
Response time10–15 seconds to a few minutes
4
RepeatabilityCantilever motion is measured without contact by a laser interferometer.
5
<1% at calibration concentrationCell / volume

Why is a small gold-coated cell important?

HF readily adheres to surfaces and reacts with moisture. Reducing volume and selecting appropriate wetted materials shorten stabilisation time and improve measurement reliability.

<0,5 ppbstandard LoD
0,2 ppb Sub-ppb measurement capability
0–10 ppmdynamic range
12 monthsrecalibration
Application data

Linear response even for reactive HF

Manufacturer tests confirmed linearity up to 400 ppb and stable detection at approximately 0.2 ppb with one-minute averaging in the higher-laser-power configuration.

400 ppbexample calibration concentration
0.2 ppbdemonstrated low-level measurement
10–20 mintypical stabilisation after gas change
<0.1 ppbobserved baseline level

Confirmed HF response linearity

GASERA ONE HF linearity chart

Stepwise HF concentrations and a comparison of set and measured values. The manufacturer study confirms a linear response up to 400 ppb.

Source: Gasera ONE HF White Paper, linearity plot for the standard-laser configuration.

Technical data

Key GASERA ONE HF specifications

The final sample path, ranges, response time and outputs must be confirmed when selecting the instrument for a specific application.

ParameterInformation
Measured componentHydrogen fluoride (HF); additional H₂O measurement is used, among other functions, for laser-frequency stabilisation
Detection limitBelow 0.5 ppb; optional configuration approximately 0.2 ppb
Measurement range0–10 ppm
Response timeConfigurable from tens of seconds to several minutes, accounting for HF adsorption and the sampling system
Measurement principleInfrared photoacoustic spectroscopy with an optical MEMS cantilever sensor
Light sourceTunable diode laser
Cell stabilisationUp to 50 °C and approximately 300 mbar
Internal volumeApproximately 30 ml
Sample flowApproximately 0.6 l/min during gas exchange
RecalibrationRecommended interval: 12 months
Housing19-inch 3U, benchtop or rack mounting
Dimensions / weight48.4 × 13.9 × 44 cm; approximately 13 kg
Operating conditions0–40 °C, non-condensing humidity, IP20
CommunicationsEthernet, USB, MODBUS and AK protocol; optional analogue outputs

Planning hydrogen fluoride monitoring?

Send us information about the HF source, expected concentrations, humidity, temperature, sampling-line length and required response time. MES-EKO will help select a complete measurement system.

Contact MES-EKO
MES-EKO Sp. z o.o.
Fast response to concentration spikes
tel. (+48) 71 348 64 02 · meseko@meseko.pl · www.meseko.pl
GASERA ONE HF — product view 5