Laboratory, installation or 24/7 monitoring
The self-contained unit with integrated pump, computer and display can operate on a laboratory bench or in a standard 19-inch instrument rack.

A configurable multi-gas trace-gas analyser combining cantilever-enhanced photoacoustic detection with selectable infrared light sources. The configuration is selected for the target gases, measurement ranges and interferences, enabling selective, simultaneous trace-gas measurements in laboratories, industry and environmental monitoring.

GASERA ONE PULSE can be configured for specific gases, interferences and sample conditions. The applications below are presented in the manufacturer materials.
Measurements of CH₄, CO₂, N₂O and fluorinated gases in air, soil and water.
Emissions from animals, livestock buildings, soil and fertilisation.
Leak detection in switchgear, equipment manufacturing and service.
Monitoring of fluranes and N₂O in operating rooms and medical installations.
Leak detection of R134a, HFO-1234yf and other HVAC/R refrigerants.
Air-exchange, tightness and flow studies using SF₆ or other tracer gases.
Evaluation of reactors, materials and research processes.
Leak measurements and performance testing of laboratory extraction systems using SF₆.
Monitoring of toxic gases, leaks and process emissions.
Continuous monitoring of chemical hazards in infrastructure and high-risk facilities.
The optional MULTIPOINT SAMPLER enables automatic sampling from up to 12 points.
Flexible configuration for development, validation and comparative measurements.
GASERA ONE PULSE combines a broadband infrared source, narrow-band optical filters and a patented optically read cantilever sensor. This allows one platform to be configured for many different gases.
This is not conventional NDIR. PULSE uses chemometric fitting of sample responses to calibration data, improving selectivity in multi-component mixtures.
GASERA ONE PULSE normally operates as a benchtop analyser or in a 19-inch rack. For field measurements it can be installed in a dedicated impact-resistant rack case.
The self-contained unit with integrated pump, computer and display can operate on a laboratory bench or in a standard 19-inch instrument rack.

The optional 19-inch 3U transport enclosure made from impact-resistant plastic protects the analyser during transport and operation outside the laboratory.




The list shows example components available in GASERA ONE PULSE configurations. The final set of gases and filters is selected for the application, ranges and possible interferences.
Detection limits declared in the manufacturer table; LoD values are given in ppm. 0.005 ppm corresponds to 5 ppb.
Manufacturer materials position PULSE as faster and simpler to operate than GC and as more stable and sensitive in many applications than conventional FTIR.
Summary of product data. Detailed values and measured components depend on the selected GASERA ONE PULSE configuration.
| Parameter | Information |
|---|---|
| Instrument type | Configurable GASERA ONE PULSE multi-gas analyser |
| Measuring principle | Photoacoustic Spectroscopy / Cantilever-Enhanced PAS |
| Light source | Broadband IR source with narrow-band optical filters |
| Gas configuration | Individually selected for target gases, ranges and sample interferences |
| Typical LoD | From ppb to sub-ppm level, depending on gas |
| Dynamic range | Typically 5 orders of magnitude, approximately 100,000 × detection limit |
| Response time | Configurable from 15 seconds to several minutes |
| Recalibration | Recommended interval of 12 months |
| Communication | Ethernet / AK protocol, optional 4–20 mA analogue outputs |
| Portable version | Optional 19-inch 3U enclosure; approximately 18 kg with analyser; external power required |
| GASERA ONE dimensions / weight | 48 × 14 × 45 cm, approximately 13 kg depending on model |
Note: filter configuration, ranges and interferences must be confirmed for the specific application.
Send the gas list, required ranges, sample matrix and operating conditions. MES-EKO will help select the GASERA ONE PULSE analyser configuration.