How to select a biogas analyser
Biogas analyser selection starts with the measurement objective, sample composition and sampling location—not with a comparison of datasheets alone.
Fermentation, upgrading and biomethane-quality control involve similar gases but different ranges, accuracy requirements and process dynamics. An instrument used for routine digester control is not automatically the right choice for gas-quality confirmation before grid injection.
1. Define the measurement objective
Decide whether the result will support process control, engine or catalyst protection, upgrading control, methane-loss balancing or biomethane-quality confirmation. This determines compounds, ranges, measurement frequency and data availability.
Process control often prioritises trends and rapid change detection. Gas-quality applications may place more emphasis on accuracy, stability and selectivity at high CH₄ levels.
| Compound | Process information | Typical relevance |
|---|---|---|
| CH₄ | Fuel fraction and process yield | Digestion, upgrading and biomethane quality |
| CO₂ | Gas dilution and CO₂-removal performance | Fermentation and upgrading |
| H₂S | Corrosion risk and desulphurisation load | Engine, catalyst and installation protection |
| O₂ | Air ingress, leakage or process condition | Safety and diagnostics |
| H₂ | Additional information in selected processes | Fermentation and research |
2. Define ranges and sampling points
The same gas may be measured before desulphurisation, after upgrading and before the final consumer. H₂S ranges can differ greatly between these points, so one range may not provide useful resolution everywhere.
Several locations can use separate analysers or a sequential multipoint system. Purge time, transport delay and the maximum acceptable interval between readings must then be considered.
3. Account for moisture and sample conditioning
Biogas is usually wet and may contain aerosols, dust and aggressive compounds. Condensation changes the sample and can damage the system. Sampling location, tubing, filtration, heating or cooling and condensate removal must be designed together with the analyser.
Drying should not be automatic. Conditioning must match the method and objective so that target compounds are not removed before measurement.
4. Select the method and operating mode
NDIR is practical for continuous measurement of infrared-active compounds such as CH₄ and CO₂. TDLAS can provide fast, selective measurement of a defined compound and range. Electrochemical sensors are often used for O₂ and H₂S, with their ranges, lifetime and interferences taken into account.
A stationary online analyser provides the trend required for control. A portable instrument is useful for diagnostics and point comparison, but periodic checks may miss short process events.
Information to prepare
- Measurement objective and use of the data
- CH₄, CO₂, H₂S, O₂ and additional compounds
- Range for every gas at every point
- Humidity, temperature, pressure and condensation risk
- Continuous, periodic or sequential measurement
- Interfaces, alarms and control-system integration
Frequently asked questions
Can one analyser cover the entire biogas process?
It can serve several points if ranges and cycle time are suitable. Very different concentrations or simultaneous data may require separate paths or analysers.
Can a portable analyser replace continuous measurement?
Not always. It is useful for periodic checks and diagnostics but cannot show the full process dynamics between visits.
Is a gas cooler always required?
No. Conditioning depends on humidity, method and target compounds. The correct solution may be a heated path, cooling or measurement without either approach.