How to design a multipoint gas measurement system
One analyser can serve many points if cycle timing, sample lines and conditioning preserve the required process information.
A multipoint system reduces instrument cost and keeps measurement consistent, but each location is only measured periodically. Transport delay, purge time and path comparability are critical.
1. Start with required data frequency
An analyser serving 12 points does not measure each continuously. The cycle includes transport, purge, stabilisation, reading and valve switching.
Fast processes may need fewer points per analyser, parallel analysers or continuous measurement at critical locations.
| Cycle element | Why it is needed | If too short |
|---|---|---|
| Transport | Moves the new sample | Result still belongs to the previous point |
| Purge | Removes residue from line and valves | Samples are mixed |
| Stabilisation | Allows analyser response to settle | Low or unstable result |
| Measurement | Records a representative value | Insufficient data |
| Switching | Changes point without disturbance | Pressure spikes and artefacts |
2. Treat tubing as part of the measurement
Length, diameter and material affect delay and losses. Reactive gases such as NH₃, HF and HCl need particular care. Wet samples can condense in colder sections.
Use similar path geometry where possible or determine transport times individually. Tightness, filtration and flow must be controlled for every line.
3. Separate common and local conditioning
Some equipment can be shared before the analyser; other elements must be installed near sampling points. A long section before filtration or heating can alter the sample.
Different pressure, temperature or dust loads should be equalised without uncontrolled dilution.
4. Preserve point identity and data quality
Every result needs point number, sampling start, stable-reading time and flow alarms. The transition period after switching should be excluded.
Plan a zero or control point, leak-test procedure and manual line selection for diagnostics.
Example cycle-time calculation
Assume 12 points and a combined two minutes per line for transport, purge, stabilisation and recording. A complete cycle then takes about 24 minutes, so each point is updated no more frequently than once per cycle. If the process can change significantly within five minutes, the presence of 12 inlets does not meet the information requirement.
Options include shorter lines, higher flow within analyser limits, splitting points between two instruments or keeping critical locations on continuous measurement. Design therefore begins with required data frequency, not the maximum number of valves.
Information to prepare
- Maximum interval between results from each point
- Length, diameter and material of each line
- Transport, purge and stabilisation times
- Pressure, temperature and humidity differences
- Local and common filtration or heating
- Logging point number, alarms and transition periods
Frequently asked questions
How many points can one analyser serve?
It depends on the instrument and required data frequency. The number of physical inlets does not guarantee adequate reporting frequency.
Must all sample lines have the same length?
Not always, but differences must be reflected in transport and purge timing. Similar geometry simplifies control and comparison.
Can reactive gases be measured with multipoint sampling?
Yes, but material, temperature, path length and cycle time must limit adsorption and losses.