TOC, THC, MHC and NMHC – what is the difference and how to select an FID analyser?
Hydrocarbon emission measurements use several similar terms: TOC, THC, MHC and NMHC. They do not always describe the same parameter. The right analyser configuration depends on the measurement objective, the required method and how the individual fractions are separated.
FID is one of the main technologies used for continuous and periodic measurement of organic compounds in gases. This guide explains the common abbreviations and the key considerations when selecting an analyser.
What does an FID detector measure?
A flame-ionisation detector (FID) measures the response of organic compounds in a hydrogen flame. Its signal is related to the organic carbon content of the sample, allowing concentration changes to be followed continuously in emissions and processes.
It provides a total result rather than a list of identified substances. Different compounds can produce different responses, so the reference gas, conversion and definition of the reported parameter matter. Separating the methane and non-methane fractions requires an appropriate configuration.
TOC, THC, MHC and NMHC — what do they mean?
| Parameter | What does it describe? | When is it needed? | Practical difference |
|---|---|---|---|
| TOC | Total Organic Carbon; a result related to the organic carbon content of the gas. | When a total organic emission result must be reported as carbon, for example in mgC/Nm³. | Describes organic carbon and the reporting basis; it should not automatically be treated as equivalent to THC. |
| THC | Total hydrocarbons, including methane. | When combined information about hydrocarbons is needed according to the selected method. | Does not exclude the methane fraction as NMHC does. |
| MHC | The methane fraction. OneFID materials use this designation for methane (CH₄) measurement. | When methane must be determined and the MHC/NMHC fractions distinguished. | Separates out methane; it is not the total result for the remaining hydrocarbons. |
| NMHC | Non-methane hydrocarbons. | When hydrocarbons other than methane must be distinguished from the methane fraction. | Excludes methane; a THC result alone does not provide this distinction. |
The exact scope of the result, units and conversion depend on the method and configuration. OneFID materials confirm that MHC represents the methane fraction; no additional expansion of the abbreviation is assumed here.
Why are THC and NMHC different?
THC covers total hydrocarbons, including methane. NMHC refers to the non-methane fraction. When methane is present, the total result does not show how much of the signal comes from the other hydrocarbons.
NMHC determination therefore requires a configuration that distinguishes methane from the remaining fraction, rather than just a standard FID total result. Results expressed in different units or on different conversion bases should not be subtracted.
When is NMHC measurement needed?
- Industrial emission monitoring when the measurement task concerns the non-methane fraction.
- Measurements at emission sources where methane must be distinguished from other hydrocarbons.
- Periodic campaigns and measurement-laboratory work when NMHC is the required parameter.
The installation type alone does not determine whether NMHC measurement is mandatory. The required parameter and method should be agreed for the specific application.
What matters when selecting an FID analyser?
- Parameter and reporting basis: Define whether TOC, THC, MHC or NMHC is required, in which units and according to which method.
- Measurement range: Specify expected concentrations and possible variations. Check the certified range as well if the application requires it.
- Sample preparation: Consider temperature, humidity, filtration, sampling location and the sample matrix. Conditioning must suit the parameter being measured.
- Line and sample-path temperature: Select heating and path length for the sample conditions and method. OneFID can use a probe with a line heated to 180°C; this is not a universal temperature for every FID application.
- Response time: Assess the required result time together with the line and sampling system, rather than the detector alone.
- Portable or stationary operation: Consider the number of points, access to power, operating gases and day-to-day handling.
- Data recording and communication: Define the required result storage, data export and communication with the customer’s system.
- Method and certification: If the application requires documented measurement performance, check the relevant certificates and their scope.
Portable FID analysers
For measurements at different points in an installation or periodic emission campaigns, a portable FID analyser can be a practical option.
Pollution OneFID
Pollution OneFID is designed for on-site TOC measurement and simultaneous MHC and NMHC determination using a single FID detector. An integrated catalyst and proprietary injection system enable these fractions to be distinguished.
- Weight of 11 kg with power supply; battery operation available.
- Response time of 30 s with a 3 m line.
- Internal flash memory, USB, Wi-Fi and Modbus TCP/IP.
How to select the right configuration
The abbreviation TOC, THC or NMHC alone is not enough to select an analyser correctly. Expected concentration ranges, sample characteristics, sampling arrangements and measurement-method requirements are also needed.
Describe your applicationFrequently asked questions
Do TOC and THC mean the same thing?
They should not be used interchangeably. TOC refers to organic carbon, while THC refers to total hydrocarbons. Check the method definition, units and conversion basis.
How does NMHC differ from THC?
NMHC describes non-methane hydrocarbons. THC includes methane, so a THC result does not replace NMHC measurement.
Does FID measure methane?
Methane produces an FID response. A separate methane result requires an appropriate configuration; OneFID identifies it as MHC.
When is NMHC measurement needed?
When the measurement objective or method requires methane to be distinguished from other hydrocarbons, for example at an emission source. Not every application requires this separation.
Can an FID analyser be portable?
Yes. OneFID is a portable analyser for on-site measurements. Selecting a complete set also involves the probe, line, power supply and operating gases.
How should the measurement range be selected?
Use expected concentrations and their variations, required sensitivity and the method. Check the available instrument range and certified range separately.