VOC, TOC, THC and NMHC — what is the difference?
Similar abbreviations may describe a compound group, a total analyser signal or a reporting convention. They should not be used interchangeably.
Before selecting an analyser, define whether the result should be total, separate methane from other hydrocarbons or identify individual compounds. This determines the method.
Essential terms
VOC means volatile organic compounds. TVOC describes a total VOC content or signal according to a defined method.
TOC refers to total organic carbon reported as carbon. THC is total hydrocarbons, MHC the methane fraction and NMHC non-methane hydrocarbons. The exact result depends on the standard, configuration and conversion.
| Term | Practical meaning | Question before measurement |
|---|---|---|
| VOC | Group of volatile organic compounds | Which compounds are included? |
| TVOC | Total VOC result according to a method | How are the total and reference defined? |
| TOC | Total organic carbon | Should the result be reported as carbon mass? |
| THC | Total hydrocarbons | Is methane included? |
| MHC | Methane hydrocarbons | How is the methane fraction determined? |
| NMHC | Non-methane hydrocarbons | Must CH₄ be separated? |
FID — a fast total result
A flame-ionisation detector measures the response of organic compounds in a hydrogen flame. It is used for continuous or portable total measurement in emissions and processes.
FID does not automatically identify every compound. Response varies between substances, so reference gas, response factors and reporting matter. MHC/NMHC separation needs an appropriate configuration.
GC or microGC — when speciation is required
Chromatography separates a mixture over time, enabling individual compounds or groups to be determined. It is appropriate when benzene, toluene or specific biomethane interferents must be identified.
Selectivity comes with a defined analysis cycle and configuration of columns, detector and operating gases. It does not replace FID in every fast total measurement.
How to write the requirement
Instead of requesting only a “VOC analyser”, define the medium, compound list or total quantity, range, unit, result time, sample temperature and humidity and reference method. This prevents comparison of instruments that report technically different quantities.
Example: two requests described as “VOC measurement”
One customer needs total organic stack emissions reported as mgC/Nm³. This points toward a heated sample path and an FID configured for TOC or THC. Another customer needs to know whether specific odorant-masking compounds are present in biomethane. A total result is insufficient; a separation method such as microGC is required.
Both requests use the term VOC but concern different measurands, analysis times and configurations. The compound list, reporting convention and reference method should be agreed before quotation.
Information to prepare
- Total result or compound identification
- TOC, THC, MHC, NMHC or TVOC
- Unit and reference substance
- Range and required result time
- Sample temperature and heated-path requirement
- Reference method or standard
Frequently asked questions
Does VOC always mean the same compound list?
No. It is a broad group, and the exact scope depends on the definition and measurement method.
Will FID identify the compound in a sample?
A standard FID result is total. Individual identification requires a separation method such as chromatography.
Are TOC and THC always equal?
This should not be assumed. Definitions, conversion and analyser response can differ, so the reporting requirement must be explicit.
Manufacturer source materials
Related solutions
Open related solution and instrument pages.
Discuss your measurement with MES-EKO
Send the gas, range, sample conditions and operating mode. We will select the method, analyser and measurement path.
Send an enquiry