One detector
A patented injection system enables simultaneous TOC, MHC and NMHC determination using one FID detector.
A complete portable FID analyser for field TOC measurements and the simultaneous determination of methanic (MHC) and non-methanic (NMHC) hydrocarbons in stack gas.
ONEFIDThe compact and robust design integrates the analyser, power and measurement media, reducing the number of separate components carried to the sampling platform.

A patented injection system enables simultaneous TOC, MHC and NMHC determination using one FID detector.
Battery, hydrogen, zero and span gas, combustion-air filter and catalyst are integrated into the instrument.
The adjustable display and interface can be used conveniently with the analyser placed vertically or horizontally.
The instrument interface can be replicated over Wi-Fi on an external device for live data and remote control.
The proprietary injection system separates the measurement tasks while retaining one detector, reducing media consumption, operating cost, weight and system size.
Total gaseous organic carbon measurement by FID in stack emissions.
Automatic determination of the methane fraction within the same measurement cycle.
Determination of non-methane hydrocarbons without adding a separate detector to the field system.
A large 7-inch touchscreen, continuous diagnostics and remote access simplify field campaigns.
Wi-Fi connectivity replicates the OneFID graphical interface on a phone, tablet or computer. The operator can view live data and control the analyser remotely.
Wi-Fi


The large 7-inch touchscreen displays operating status, the active method and current concentrations. The interface remains usable in either vertical or horizontal orientation.

The instrument combines high portability with the requirements of stack-emission measurement and hydrocarbon-fraction analysis.
On-site TOC concentration analysis by FID in accordance with EN 12619:2013.
Automatic methane determination in accordance with EN ISO 25140:2010 and determination of the NMHC fraction.
Portable survey campaigns in industrial and environmental applications.
OneFID is designed for stack-emission measurements that require documented measurement performance.
OneFID combines a certified measurement configuration with an 11 kg system weight, heating time below 20 minutes and battery operation.
Gas storage, hydrogen and accessories can be integrated into the set, while external cylinders can be connected through quick couplings.
Analyser, sample path, power, auxiliary-gas and communication parameters.
| Complete technical features | |
|---|---|
| Size and weight | 403 × 213 × 433 mm; 11 kg with power supply |
| Power source | external power adapter 110-240 VAC / 24 VDC / 300 W; 1 battery |
| Environmental operating conditions | 0°C < T < +50°C; 0% < RH < 95% |
| Detector body maximum temperature | 185°C |
| Sample inlet maximum temperature | 185°C |
| Aspirated sample flow | 900 ml/min at the inlet fitting |
| Sampling-point pressure | atmospheric ±100 mbar |
| Measuring ranges | measuring range 0-5,000 mgC/Nm³; certified range 0-15 mgC/Nm³; supplementary ranges 0-500, 0-1,000 and 0-2,000 mgC/Nm³ |
| Accuracy | the greater of 1% of full scale or 0.4 mgC/Nm³ (0.25 ppm as propane) |
| Limit of detection | <0.1 mgC/Nm³ |
| Response time | 30 s with a 3 m line |
| Method management and data storage | upload and download methods via the user interface; internal flash memory |
| Interfaces | USB for data download; Wi-Fi for connection from an external device, data visualisation and remote control; Modbus TCP/IP |
| Flame management | electronic management with diagnostics, visual flame-out alarm and automatic restart |
| Hydrogen storage cartridge | metal-hydride technology; 20 h continuous operation; rechargeable from an external pressurised hydrogen source; no restrictions on transport by car or plane in accordance with UN 3479 |
| Zero air | non-refillable 1 L technical-air can at 12 bar / 20°C; 10 h continuous operation; quick connector for an external zero-air cylinder; configurable front-panel connectors; shipping: UN 1956, hazard class 2 division 2.2 - non-flammable, non-toxic gases |
| Span gas | non-refillable 1 L propane-in-air can at 12 bar / 20°C; total-carbon-equivalent concentrations 16, 40, 120 or 320 mgC/Nm³; other configurations on request; quick connector for an external span-gas cylinder; configurable front-panel connectors; shipping: UN 1956, hazard class 2 division 2.2 - non-flammable, non-toxic gases |
| Certifications | EN 15267-4; EN 12619:2013; QAL1 quality requirements specified in EN 14181 |
The adjustable display allows the analyser position to be matched to the space available on the sampling platform.


Components can be selected for the operating method, required autonomy and sampling-point configuration.

12 cans: Span Gas 16/40/120/320 mgC/Nm³, Zero Air or a 9 Zero + 3 Span set.

Metal hydrides and rare-earth technology; rechargeable from a pure-hydrogen source.
50 l · 200 × 30 mm
Activated-carbon filter for cleaning the detector combustion air.
220 × 28 mm
Powers the instrument, probe and line; charged without removal from the analyser.

EN 12619:2013-compliant probe with a self-regulating heated sample line.
Send the plant type, expected TOC range, stack-gas temperature and humidity, and required sample-line length.