POLLUTION · PORTABLE STACK-EMISSIONS ANALYSER

OneFID – portable FID analyser for TOC and NMHC

A complete portable FID analyser for field TOC measurements and the simultaneous determination of methanic (MHC) and non-methanic (NMHC) hydrocarbons in stack gas.

Pollution OneFID portable analyserONEFID
0-5,000mgC/Nm³ overall range
<0.1mgC/Nm³ detection limit
30 sresponse with a 3 m line
11 kgwith power supply
<20 minheating time
20 hH₂ cartridge operation
One instrument - complete measurement

A portable FID ready for field operation

The compact and robust design integrates the analyser, power and measurement media, reducing the number of separate components carried to the sampling platform.

OneFID in horizontal position
1

One detector

A patented injection system enables simultaneous TOC, MHC and NMHC determination using one FID detector.

✓

All-in-One

Battery, hydrogen, zero and span gas, combustion-air filter and catalyst are integrated into the instrument.

↕

Two operating positions

The adjustable display and interface can be used conveniently with the analyser placed vertically or horizontally.

↔

SMART connectivity

The instrument interface can be replicated over Wi-Fi on an external device for live data and remote control.

Measurement technology

TOC, MHC and NMHC without a second detector

The proprietary injection system separates the measurement tasks while retaining one detector, reducing media consumption, operating cost, weight and system size.

TOC

Total gaseous organic carbon measurement by FID in stack emissions.

MHC / CH₄

Automatic determination of the methane fraction within the same measurement cycle.

NMHC

Determination of non-methane hydrocarbons without adding a separate detector to the field system.

Operation and automation

The operator sees the result while the instrument manages the process

A large 7-inch touchscreen, continuous diagnostics and remote access simplify field campaigns.

Integrated control software

  • gas-can pressure control,
  • detector and catalyst temperature control,
  • flame monitoring and automatic restart,
  • sample-flow control,
  • automatic calibration procedures,
  • diagnostic checks and visual alarms.

Data and communication

  • real-time result display,
  • interface replication on an external device over Wi-Fi,
  • remote analyser control,
  • USB data download,
  • Modbus TCP/IP,
  • internal flash memory and method upload/download.
SMART connectivity

The same interface on the analyser and an external device

Wi-Fi connectivity replicates the OneFID graphical interface on a phone, tablet or computer. The operator can view live data and control the analyser remotely.

OneFID connected wirelessly to external devicesWi-Fi
OneFID display on a mobile device
Mobile charts and measurement results
OneFID measurement chart over Wi-Fi
Real-time data visualisation
Remote OneFID operation from a computer
Interface replication and remote control
Measurement display

TOC, MHC and NMHC shown together

The large 7-inch touchscreen displays operating status, the active method and current concentrations. The interface remains usable in either vertical or horizontal orientation.

  • immediate on-site assessment,
  • clear flame and alarm status,
  • methods, calibration and diagnostics in one place.
OneFID measurement screen with THC, MHC and NMHC results
Applications

OneFID in field measurements

The instrument combines high portability with the requirements of stack-emission measurement and hydrocarbon-fraction analysis.

TOC

Stack emissions

On-site TOC concentration analysis by FID in accordance with EN 12619:2013.

CH₄

MHC and NMHC

Automatic methane determination in accordance with EN ISO 25140:2010 and determination of the NMHC fraction.

VOC

Environmental screening

Portable survey campaigns in industrial and environmental applications.

Measurement compliance

Standards and certification

OneFID is designed for stack-emission measurements that require documented measurement performance.

✓
TÜVcertification of instrument measurement performance
✓
EN 15267-4:2024requirements for portable automated measuring systems
✓
EN 12619:2013determination of total gaseous organic carbon by FID
✓
QAL1 to EN 14181quality requirements for automated measuring systems
✓
EPA Method 25Adetermination of total gaseous organic concentration

Ready for field operation

OneFID combines a certified measurement configuration with an 11 kg system weight, heating time below 20 minutes and battery operation.

Gas storage, hydrogen and accessories can be integrated into the set, while external cylinders can be connected through quick couplings.

Technical data

Complete OneFID specification

Analyser, sample path, power, auxiliary-gas and communication parameters.

Complete technical features
Size and weight403 × 213 × 433 mm; 11 kg with power supply
Power sourceexternal power adapter 110-240 VAC / 24 VDC / 300 W; 1 battery
Environmental operating conditions0°C < T < +50°C; 0% < RH < 95%
Detector body maximum temperature185°C
Sample inlet maximum temperature185°C
Aspirated sample flow900 ml/min at the inlet fitting
Sampling-point pressureatmospheric ±100 mbar
Measuring rangesmeasuring range 0-5,000 mgC/Nm³; certified range 0-15 mgC/Nm³; supplementary ranges 0-500, 0-1,000 and 0-2,000 mgC/Nm³
Accuracythe greater of 1% of full scale or 0.4 mgC/Nm³ (0.25 ppm as propane)
Limit of detection<0.1 mgC/Nm³
Response time30 s with a 3 m line
Method management and data storageupload and download methods via the user interface; internal flash memory
InterfacesUSB for data download; Wi-Fi for connection from an external device, data visualisation and remote control; Modbus TCP/IP
Flame managementelectronic management with diagnostics, visual flame-out alarm and automatic restart
Hydrogen storage cartridgemetal-hydride technology; 20 h continuous operation; rechargeable from an external pressurised hydrogen source; no restrictions on transport by car or plane in accordance with UN 3479
Zero airnon-refillable 1 L technical-air can at 12 bar / 20°C; 10 h continuous operation; quick connector for an external zero-air cylinder; configurable front-panel connectors; shipping: UN 1956, hazard class 2 division 2.2 - non-flammable, non-toxic gases
Span gasnon-refillable 1 L propane-in-air can at 12 bar / 20°C; total-carbon-equivalent concentrations 16, 40, 120 or 320 mgC/Nm³; other configurations on request; quick connector for an external span-gas cylinder; configurable front-panel connectors; shipping: UN 1956, hazard class 2 division 2.2 - non-flammable, non-toxic gases
CertificationsEN 15267-4; EN 12619:2013; QAL1 quality requirements specified in EN 14181
Sample path: an EN 12619:2013-compliant probe is available with a specially designed sample line self-regulated at 180°C.
Ergonomics

One instrument, two positions

The adjustable display allows the analyser position to be matched to the space available on the sampling platform.

Accessories and consumables

Complete field system

Components can be selected for the operating method, required autonomy and sampling-point configuration.

OneFID calibration gases

Calibration gases

12 cans: Span Gas 16/40/120/320 mgC/Nm³, Zero Air or a 9 Zero + 3 Span set.

OneFID H2 cartridge

H₂ cartridge

Metal hydrides and rare-earth technology; rechargeable from a pure-hydrogen source.

50 l · 200 × 30 mm
OneFID carbon filter

Carbon filter

Activated-carbon filter for cleaning the detector combustion air.

220 × 28 mm
OneFID battery

Battery

Powers the instrument, probe and line; charged without removal from the analyser.

OneFID sampling probe

Probe and 180°C line

EN 12619:2013-compliant probe with a self-regulating heated sample line.

Also available: external battery charger, carrying case, spare hydrogen cartridge and filters.

Select a OneFID system

Send the plant type, expected TOC range, stack-gas temperature and humidity, and required sample-line length.