POLLUTION · Quality & Process

PicoGC VOC

An automatic micro gas chromatograph for continuous VOC monitoring in biogas-upgrading plants. It verifies activated-carbon filter performance, detects breakthrough and protects the quality and odorability of biomethane intended for grid injection.

Pollution PicoGC VOC explosion-proof analyserPicoGC VOC
10 VOCsconfigured compounds
0-100 ppmrange per compound
1 ppmlimit of detection
<300 sanalysis time
-35...+60°Coutdoor operation
IP66PicoGC and RCB enclosures
Process challenge

VOCs reveal when an activated-carbon filter stops protecting the plant

Raw biogas VOC concentrations can reach 5,000 ppm. If they are not removed effectively, VOCs may clog the upgrading system and interfere with biomethane odorisation.

Fresh carbon

The activated-carbon filter reduces VOCs to below 5 ppm and the analyser confirms correct performance.

Increasing saturation

PicoGC alternates between two downstream sample points and tracks changes in marker compounds.

Breakthrough detected

A rapid VOC increase triggers an alarm and identifies the filter that needs replacement.

Typical total VOC increase during activated-carbon filter breakthrough

Typical trend: VOC concentration increases rapidly after carbon saturation. Continuous monitoring uses the full carbon charge without compromising protection of the upgrading plant.

Analysed compounds

Ketones, BTEX and terpenes in one system

Each listed compound has a 0-100 ppm range and a stated 1 ppm detection limit. Other VOCs may be available on request.

Acetoneketone · C3H6O
2-butanone (MEK)ketone · C4H8O
BenzeneBTEX · C6H6
TolueneBTEX · C7H8
EthylbenzeneBTEX · C8H10
Xyleneso-, p- and m-xylene
Pinenealpha and beta
o,p-cymenearomatic terpene
Limoneneterpene · C10H16
3-careneterpene · C10H16
Measurement system

PicoGC at the process line, RCB at the plant control system

The analyser operates close to two sample points in the ATEX area, while the Remote Control Box stores, processes and transfers data to the control room or Pollution Guardian cloud.

PicoGC

  • microGC in an ATEX Zone 1 enclosure,
  • two process-gas inlets,
  • automatic sample-point switching,
  • chromatogram acquisition and storage,
  • configurable lower and upper alarms.

RCB and communication

  • Ethernet Modbus TCP/IP and Web APIs,
  • optional RS485 Modbus ASCII/RTU,
  • Wi-Fi and 4G,
  • PLC/DCS/SCADA integration,
  • remote diagnostics and Pollution Guardian.
Pollution PicoGC analyser production
Field-ready construction

The proven PicoGC platform

The VOC version shares the robust mechanical architecture of PicoGC THT-TBM, but uses its own chromatographic method, two sample inlets and a configuration dedicated to upgrading-plant protection.

  • GC enclosure below 27 kg,
  • low carrier-gas consumption,
  • automatic restart after power interruption,
  • more than three years of chromatogram storage.
Interior and utilities

A compact microGC for continuous operation

Carrier gas is grade 5.5 helium or hydrogen. At four analyses per hour, a 14 L / 200 bar cylinder can last up to 10 or 9 years respectively.

PicoGC VOC enclosure interior

Two sampling points

Two 1/8-inch connections enable alternating supervision of two streams downstream of the carbon filters.

Sample conditions

Recommended 20-100 mbarg, maximum 1.5 barg; consumption 13 mL/min at 20 mbar; H2O below 2,000 ppm and particles below 2 µm.

Low operator workload

Automatic measurement, alarming and data transfer reduce manual sampling and premature carbon replacement.

Technical data

PicoGC VOC specifications

Key analyser, sample, carrier-gas, power and communication parameters.

Complete technical specifications
Applicationautomatic gas chromatograph for VOC monitoring
Functionschromatogram acquisition, alarm management and continuous monitoring; analogue and digital I/O status, data protection, PLC communication and supervision, remote server and wireless connectivity
Inputs/Outputs - RCB1 × RS485 Modbus ASCII/RTU (optional); 1 × Ethernet TCP/IP Modbus / Web APIs; Wi-Fi; 4G
Inputs/Outputs - GC1 × power supply and data RJ45 (Ethernet)
Supported protocolsModbus TCP/IP; Web APIs; Modbus RTU (optional)
ATEX GC enclosureIP66; 450 × 350 × 251 mm; below 27 kg
CertificationsCE ATEX II 2G Ex db IIB+H2 T6 Gb; IECEx Ex db IIB+H2 T6 Gb
RCB enclosureIP66; 325 × 428.2 × 178 mm; 4 kg
Carrier gas connection1 × carrier-gas inlet, 1/8-inch OD Swagelok type
Process gas connections2 × sample-gas inlet, 1/8-inch OD Swagelok type
Carrier gashelium (He) or hydrogen (H2); 14 L / 200 barg He cylinder: up to 10 years at 4 analyses/hour; 14 L / 200 barg H2 cylinder: up to 9 years at 4 analyses/hour; minimum purity Grade 5.5 (≥99.9995%); pressure 3.5 ±0.5 barg
Sample gas conditionsrecommended pressure 20-100 mbarg; maximum 1.5 barg; consumption 13 ml/min at 20 mbar (excluding fast-loop bypass, if required); free of particulates and liquids; H2O <2000 ppm; particulates Ø <2 µm
Analysed compoundRangeLimit of detection (S/N = 3)
Acetone0-100 ppmLOD 1 ppm
2-Butanone - MEK0-100 ppmLOD 1 ppm
Benzene0-100 ppmLOD 1 ppm
Toluene0-100 ppmLOD 1 ppm
Ethylbenzene0-100 ppmLOD 1 ppm
Xylenes (o-, p-, m-)0-100 ppmLOD 1 ppm
Pinene (alpha and beta)0-100 ppmLOD 1 ppm
o,p-Cymene0-100 ppmLOD 1 ppm
Limonene0-100 ppmLOD 1 ppm
3-Carene0-100 ppmLOD 1 ppm
Data loggingconcentration storage with configurable lower and upper alarms
Chromatogram storagemore than 3 years of continuous measurements with configurable lower and upper alarms
LanguageEnglish
Operating temperature-35°C to +60°C; outdoor installation
Power supply24 V supplied by external RCB, included; automatic restart after power interruption
Nominal power consumption35 W at 0°C; maximum instantaneous power draw 90 W
Analysis timebelow 300 seconds
Configuration: the final method and compound set should be confirmed for the biogas composition, upgrading technology and expected interferences.

Select PicoGC VOC for your plant

Provide the biogas composition, upgrading technology, carbon-filter arrangement, sampling points and required automation interface.

meseko@meseko.pl
(+48) 71 348 64 02