POLLUTION · gas and energy

PicoGC THT‑TBM

An automatic micro gas chromatograph for continuous, unattended determination of THT or TBM in natural gas, biomethane and LPG. Its ATEX construction permits installation directly in the hazardous area.

Pollution PicoGC THT-TBM explosion-proof analyserPicoGC THT-TBM
0–200mg/Sm³ · THT/TBM range
0.5 ppmdetection limit
30–120 sanalysis time
−35…+60°Coutdoor operation
IP66field enclosure
<1 L/yeargas release with reinjection
Measurement system

Online results without laboratory sampling

The system combines the explosion-proof PicoGC analyser with an RCB remote control box. A result is available after every cycle, while configurable alarms flag concentration limits.

PicoGC analyser interior

PicoGC in Zone 1

The chromatographic module, pneumatics and electronics operate inside an Ex db enclosure. Automatic restart limits gaps after a power interruption.

RCB outside the zone

The Remote Control Box powers the analyser and provides Ethernet, RS‑485, Wi‑Fi and 4G connectivity.

One configuration per odorant

The chromatographic column is selected specifically for THT or TBM. The TBM method addresses the mercaptan mixture commonly used for odorisation.

microGC technology

Fast separation and selective odorant detection

A precision micro-injector transfers the sample to an odorant-specific column. The μTCD detects the separated THT or TBM peak and the software calculates concentration automatically.

MEMS

Miniaturised injector and detector components provide low internal volume and a short analytical cycle.

Carrier gas

Helium or hydrogen, purity grade 5.5. A 14 L cylinder at 200 bar supports years of operation at four analyses per hour.

Diagnostics

The system supervises operating parameters, stores chromatograms and supports remote service actions.

Connectivity and data

Results in the control room, SCADA and cloud

PicoGC stores more than three years of results and chromatograms. Data can be displayed locally or transferred to higher-level systems.

Plant integration

  • Ethernet and Modbus TCP/IP,
  • optional RS‑485 with Modbus ASCII/RTU,
  • Web APIs for data integration,
  • upper and lower concentration alarms,
  • Wi‑Fi and 4G modem.

Pollution Guardian

  • live and historical results,
  • charts, tables and statistics,
  • SMS, email and push alarms,
  • remote diagnostics and control,
  • encrypted communications, VPN and SCADA integration.
Installation points

Odorisation control across the network

Continuous measurement reveals actual concentration changes between network points and reduces the time for which an off-specification condition can remain unknown.

01

Before and after odorisation

Verification of odorant dosing at metering, pressure-control and pressure-reduction stations.

02

Critical network points

Monitoring at remote locations, hard-to-access sites and distribution-network endpoints.

03

Natural gas, biomethane and LPG

THT or TBM control in fuel gases without organising periodic sample collection and transport.

Architecture

PicoGC, RCB and supervisory system

Sample and calibration gas enter the analyser in the hazardous area, while the RCB provides power and communications. Optional reinjection returns analysed gas to the pipeline.

PicoGC system diagram with RCB, communications and sample reinjection
Chromatographic result

THT or TBM determination

Each configuration uses a method and column matched to the selected odorant.

Example THT chromatogram
THT: 0–200 mg/Sm³ range, 0.5 ppm LOD (1.86 mg/Sm³), repeatability <0.6% RSD at 32 mg/Sm³.
Example TBM chromatogram
TBM: 0–200 mg/Sm³ range, 0.5 ppm LOD (1.91 mg/Sm³), repeatability <3% RSD at 9.3 mg/Sm³.
Technical data

PicoGC THT‑TBM specifications

Analyser, sample, carrier-gas, power and communication parameters.

Complete technical features
Applicationautomated gas chromatograph for quantitative analysis of THT or TBM odorants
Functionschromatogram acquisition, alarm management and monitoring; analogue and logic I/O status, data protection, PLC communication and supervision, remote server access and wireless connectivity
Inputs/Outputs from RCB1 × RS485 dedicated to Modbus ASCII/RTU communication (optional); 1 × Ethernet TCP/IP Modbus / Web APIs; Wi-Fi; 4G
Inputs/Outputs from GC1 × power supply and RJ45 data (Ethernet)
Supported protocolsModbus RTU; Modbus TCP/IP; Web APIs
ATEX GC enclosureIP66; 450 × 350 × 251 mm; below 27 kg
CertificationsCE ATEX II 2G Ex db IIB+H2 T6 Gb; IECEx Ex db IIB+H2 T6 Gb
RCB enclosureIP66; 325 × 428.2 × 178 mm; 4 kg
Carrier gas connection1 × carrier-gas inlet, 1/8-inch OD Swagelok type
Process gas connections1 × sample-gas inlet, 1 × calibration-gas inlet and 1 × sample-gas outlet for reinjection; all 1/8-inch OD Swagelok type
Carrier gashelium (He) or hydrogen (H2); 14 L / 200 barg He cylinder: 9 years at 4 analyses/hour; 14 L / 200 barg H2 cylinder: 8 years at 4 analyses/hour; purity Grade 5.5 (≥99.9995%); pressure 3.5 ±0.5 barg
Sample gas conditionssuggested pressure 20 mbarg; maximum 1.5 barg; free of liquids and particles; H2O <2000 ppm; particles Ø <2 µm
Natural-gas release to atmospherebelow 1 L/year in the configuration with reinjection
Analysed compoundRangeDetection limit (S/N = 3)
THT0-200 mg/Sm³0.5 ppm - 1.86 mg/Sm³
TBM0-200 mg/Sm³0.5 ppm - 1.91 mg/Sm³
Repeatabilityretention time ≤0.04% RSD; THT <0.6% RSD at 32.0 mg/Sm³ (32 ±0.5 mg/Sm³); TBM <3% RSD at 9.3 mg/Sm³ (9.3 ±0.6 mg/Sm³)
Data loggingTHT or TBM concentration storage in mg/Sm³ or ppm, with lower and/or upper alarm
Chromatogram storagemore than 3 years of all analytical data, with lower and/or upper alarm
LanguageEnglish
Operating temperature-35°C to +60°C; outdoor installation
Power supply24 V provided by the external RCB remote-control and communication system, included; automatic restart after power failure
Nominal power consumption35 W at 0°C; maximum instantaneous power draw 90 W
Minimum analysis time30-120 seconds, depending on the carrier gas
Configuration: THT and TBM use different chromatographic columns. The variant must match the odorant used in the network.

Select a PicoGC configuration

Provide the odorant (THT or TBM), fuel gas, pressure, measurement location and required data interface.