PicoGC in Zone 1
The chromatographic module, pneumatics and electronics operate inside an Ex db enclosure. Automatic restart limits gaps after a power interruption.
An automatic micro gas chromatograph for continuous, unattended determination of THT or TBM in natural gas, biomethane and LPG. Its ATEX construction permits installation directly in the hazardous area.
PicoGC THT-TBMThe system combines the explosion-proof PicoGC analyser with an RCB remote control box. A result is available after every cycle, while configurable alarms flag concentration limits.

The chromatographic module, pneumatics and electronics operate inside an Ex db enclosure. Automatic restart limits gaps after a power interruption.
The Remote Control Box powers the analyser and provides Ethernet, RS‑485, Wi‑Fi and 4G connectivity.
The chromatographic column is selected specifically for THT or TBM. The TBM method addresses the mercaptan mixture commonly used for odorisation.
A precision micro-injector transfers the sample to an odorant-specific column. The μTCD detects the separated THT or TBM peak and the software calculates concentration automatically.
Miniaturised injector and detector components provide low internal volume and a short analytical cycle.
Helium or hydrogen, purity grade 5.5. A 14 L cylinder at 200 bar supports years of operation at four analyses per hour.
The system supervises operating parameters, stores chromatograms and supports remote service actions.
PicoGC stores more than three years of results and chromatograms. Data can be displayed locally or transferred to higher-level systems.
Continuous measurement reveals actual concentration changes between network points and reduces the time for which an off-specification condition can remain unknown.
Verification of odorant dosing at metering, pressure-control and pressure-reduction stations.
Monitoring at remote locations, hard-to-access sites and distribution-network endpoints.
THT or TBM control in fuel gases without organising periodic sample collection and transport.
Sample and calibration gas enter the analyser in the hazardous area, while the RCB provides power and communications. Optional reinjection returns analysed gas to the pipeline.

Each configuration uses a method and column matched to the selected odorant.


Analyser, sample, carrier-gas, power and communication parameters.
| Complete technical features | ||
|---|---|---|
| Application | automated gas chromatograph for quantitative analysis of THT or TBM odorants | |
| Functions | chromatogram acquisition, alarm management and monitoring; analogue and logic I/O status, data protection, PLC communication and supervision, remote server access and wireless connectivity | |
| Inputs/Outputs from RCB | 1 × RS485 dedicated to Modbus ASCII/RTU communication (optional); 1 × Ethernet TCP/IP Modbus / Web APIs; Wi-Fi; 4G | |
| Inputs/Outputs from GC | 1 × power supply and RJ45 data (Ethernet) | |
| Supported protocols | Modbus RTU; Modbus TCP/IP; Web APIs | |
| ATEX GC enclosure | IP66; 450 × 350 × 251 mm; below 27 kg | |
| Certifications | CE ATEX II 2G Ex db IIB+H2 T6 Gb; IECEx Ex db IIB+H2 T6 Gb | |
| RCB enclosure | IP66; 325 × 428.2 × 178 mm; 4 kg | |
| Carrier gas connection | 1 × carrier-gas inlet, 1/8-inch OD Swagelok type | |
| Process gas connections | 1 × sample-gas inlet, 1 × calibration-gas inlet and 1 × sample-gas outlet for reinjection; all 1/8-inch OD Swagelok type | |
| Carrier gas | helium (He) or hydrogen (H2); 14 L / 200 barg He cylinder: 9 years at 4 analyses/hour; 14 L / 200 barg H2 cylinder: 8 years at 4 analyses/hour; purity Grade 5.5 (≥99.9995%); pressure 3.5 ±0.5 barg | |
| Sample gas conditions | suggested pressure 20 mbarg; maximum 1.5 barg; free of liquids and particles; H2O <2000 ppm; particles Ø <2 µm | |
| Natural-gas release to atmosphere | below 1 L/year in the configuration with reinjection | |
| Analysed compound | Range | Detection limit (S/N = 3) |
|---|---|---|
| THT | 0-200 mg/Sm³ | 0.5 ppm - 1.86 mg/Sm³ |
| TBM | 0-200 mg/Sm³ | 0.5 ppm - 1.91 mg/Sm³ |
| Repeatability | retention time ≤0.04% RSD; THT <0.6% RSD at 32.0 mg/Sm³ (32 ±0.5 mg/Sm³); TBM <3% RSD at 9.3 mg/Sm³ (9.3 ±0.6 mg/Sm³) | |
|---|---|---|
| Data logging | THT or TBM concentration storage in mg/Sm³ or ppm, with lower and/or upper alarm | |
| Chromatogram storage | more than 3 years of all analytical data, with lower and/or upper alarm | |
| Language | English | |
| Operating temperature | -35°C to +60°C; outdoor installation | |
| Power supply | 24 V provided by the external RCB remote-control and communication system, included; automatic restart after power failure | |
| Nominal power consumption | 35 W at 0°C; maximum instantaneous power draw 90 W | |
| Minimum analysis time | 30-120 seconds, depending on the carrier gas | |
Provide the odorant (THT or TBM), fuel gas, pressure, measurement location and required data interface.