Wastewater treatment and biogas plants
Personal monitoring for H₂S or oxygen, depending on the hazards at the worksite.
Portable detector for personal protection. Each unit is configured for one gas. It warns the wearer using visual, audible and vibration alarms.
EASYGAS PGD100Personal monitoring for H₂S or oxygen, depending on the hazards at the worksite.
Warning of exposure to a selected gas, such as CO near combustion equipment.
Selection for a specific substance, such as NH₃ or Cl₂ where present.
Worn by staff working near a known gas hazard.
Live readings for the selected gas in the air around the wearer.

Gas reaches the electrochemical sensor and the detector displays the concentration in ppm or vol.%, depending on the version. The version is selected for the target gas before purchase.
When the concentration reaches a configured threshold, the detector alerts the wearer using LEDs, vibration and sound (approximately 100 dB at 10 cm). Low and high thresholds are available, with TWA and STEL alarms on applicable versions.

Use the rear clip to attach the detector to workwear. Wear it in the breathing zone and keep the gas inlet clear. The standard package includes the detector with clip, a gas calibration cap and a quick-start guide.
A button operates the menu. The display indicates battery status, alarms and faults; the self-test checks the sensor, circuitry and battery. Authorised staff change alarm thresholds according to site procedures.
The photos show the H₂S version. The marking on the housing depends on the gas version ordered.
These codes identify separate versions. All listed versions use an electrochemical sensor selected for the target gas.
| Gas and model code | Range | Resolution |
|---|---|---|
| O₂ · oxygenPGD100-O2 | 0–30 vol.% | 0.1 vol.% |
| O₂ · lead-free sensorPGD100-O2LF | 0–30 vol.% | 0.1 vol.% |
| CO · carbon monoxidePGD100-CO | 0–1000 ppm | 1 ppm |
| CO · extended rangePGD100-COE | 0–2000 ppm | 1 ppm |
| CO · H₂ compensationPGD100-COH | 0–500 ppm | 1 ppm |
| H₂S · hydrogen sulfidePGD100-H2S | 0–100 ppm | 0.1 ppm |
| H₂ · hydrogenPGD100-H2 | 0–1000 ppm | 1 ppm |
| SO₂ · sulfur dioxidePGD100-SO2 | 0–20 ppm | 0.1 ppm |
| Cl₂ · chlorinePGD100-Cl2 | 0–50 ppm | 0.1 ppm |
| HCl · hydrogen chloridePGD100-HCL | 0–15 ppm | 0.1 ppm |
| HCN · hydrogen cyanidePGD100-HCN | 0–100 ppm | 0.1 ppm |
| PH₃ · phosphinePGD100-PH3 | 0–20 ppm | 0.1 ppm |
| ETO · ethylene oxidePGD100-ETO | 0.1 ppm | |
| O₃ · ozonePGD100-O3 | 0–5 ppm | 0.01 ppm |
| NO · nitric oxidePGD100-NO | 0–250 ppm | 1 ppm |
| NO₂ · nitrogen dioxidePGD100-NO2 | 0–30 ppm | 0.1 ppm |
| NH₃ · ammoniaPGD100-NH3 | 0–100 ppm | 1 ppm |
| NH₃ · extended rangePGD100-NH3E | 0–1000 ppm | 1 ppm |
| Display | 1.6″ FSTN LCD |
|---|---|
| Dimensions | 66 × 69 × 24 mm without clip; 66 × 69 × 41 mm with clip |
| Weight | 105 g with clip |
| Relative humidity | 0–95% RH, non-condensing |
| Operating temperature | Most versions: −30 to +50 °C. Cl₂, O₃ and ETO: −20 to +50 °C. NH₃ and NH₃E: −40 to +50 °C. O₂LF: −40 to +60 °C. |
| Enclosure protection | IP68 |
3.6 V, 1.65 Ah lithium cell (2/3 AA). Over 2 years without battery replacement in a profile of 8 hours of use per day, with up to 5 minutes of alarm activity and BLE switched off. Actual runtime depends on use.
Arrange replacement with Tecnocontrol. Only authorised personnel should replace the battery, in a safe area and in accordance with the manufacturer’s procedure.
The instrument has a two-year warranty from purchase. Tecnocontrol lists a two-year sensor warranty for O₂, CO, H₂S, NO₂ and PH₃, and one year for H₂, NO, SO₂, Cl₂, NH₃, O₃, HCl, HCN and ETO.
The warranty period does not establish sensor life. Service life depends on the version and exposure conditions.
Before use, the manufacturer recommends a bump test with gas above the alarm threshold. It checks detector response and alarm signalling.
Calibration adjusts the readings. The manufacturer recommends gas calibration at least every 180 days, following the procedure for the specific sensor. Also follow site procedures and act on test results.
For versions listed in the certificates: ATEX KSCP 24ATEX0031X (II 1G Ex ia IIC T4 Ga) and IECEx KSCP 24.0053X (Ex ia IIC T4 Ga). Before use, verify the variant code, unit marking, special conditions X and sensor operating temperature. Do not open the enclosure in an explosive atmosphere; follow the documented electrostatic precautions.
The detector can help protect against one selected hazard. On its own, it does not provide full atmospheric monitoring before entry or during work. Entry procedures and monitoring must cover all hazards identified for the site.
We can help select the EasyGas PGD100 for the hazards at your worksite.
Tell us the target gas, installation location and alarm requirements. We can help select the device and plan the integration.